DOI 10.1016/j.surg.2014.10.015
Force-Sensing Enhanced Simulation Environment (ForSense) for laparoscopic surgery training and assessment.
Аннотация
BACKGROUND
Excessive or inappropriate tissue interaction force during laparoscopic surgery is a recognized contributor to surgical error, especially for robotic surgery. Measurement of force at the tool-tissue interface is, therefore, a clinically relevant skill assessment variable that may improve effectiveness of surgical simulation. Popular box trainer simulators lack the necessary technology to measure force. The aim of this study was to develop a force sensing unit that may be integrated easily with existing box trainer simulators and to (1) validate multiple force variables as objective measurements of laparoscopic skill, and (2) determine concurrent validity of a revised scoring metric.
METHODS
A base plate unit sensitized to a force transducer was retrofitted to a box trainer. Participants of 3 different levels of operative experience performed 5 repetitions of a peg transfer and suture task. Multiple outcome variables of force were assessed as well as a revised scoring metric that incorporated a penalty for force error.
RESULTS
Mean, maximum, and overall magnitudes of force were significantly different among the 3 levels of experience, as well as force error. Experts were found to exert the least force and fastest task completion times, and vice versa for novices. Overall magnitude of force was the variable most correlated with experience level and task completion time. The revised scoring metric had similar predictive strength for experience level compared with the standard scoring metric.
CONCLUSION
Current box trainer simulators can be adapted for enhanced objective measurements of skill involving force sensing. These outcomes are significantly influenced by level of expertise and are relevant to operative safety in laparoscopic surgery. Conventional proficiency standards that focus predominantly on task completion time may be integrated with force-based outcomes to be more accurately reflective of skill quality.