DOI 10.1111/jce.15430

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2022

A real-world experience of atrioventricular synchronous pacing with leadless ventricular pacemakers.

Kowlgi Gurukripa N, Tseng Andrew S, Tempel Nathan D, Henrich Mark J, Venkatachalam Kalpathi L, Scott Luis, Shen Win-Kuang, Deshmukh Abhishek J, Madhavan Malini, Lee Hon-Chi, Asirvatham Samuel J, Friedman Paul A, Cha Yong-Mei, Mulpuru Siva K

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: The MicraTM transcatheter pacing system (TPS) (Medtronic) is the only leadless pacemaker that promotes atrioventricular (AV) synchrony via accelerometer-based atrial sensing. Data regarding the real-world experience with this novel system are scarce. We sought to characterize patients undergoing MicraTM -AV implants, describe percentage AV synchrony achieved, and analyze the causes for suboptimal AV synchrony.

METHODS

In this retrospective cohort study, electronic medical records from 56 consecutive patients undergoing MicraTM -AV implants at the Mayo Clinic sites in Minnesota, Florida, and Arizona with a minimum follow-up of 3 months were reviewed. Demographic data, comorbidities, echocardiographic data, and clinical outcomes were compared among patients with and without atrial synchronous ventricular pacing (AsVP) ≥ 70%.

RESULTS

Sixty-five percent of patients achieved AsVP ≥ 70%. Patients with adequate AsVP had smaller body mass indices, a lower proportion of congestive heart failure, and prior cardiac surgery. Echocardiographic parameters and procedural characteristics were similar across the two groups. Active device troubleshooting was associated with higher AsVP. The likely reasons for low AsVP were small A4-wave amplitude, high ventricular pacing burden, and inadequate device reprogramming. Importantly, in patients with low AsVP, subjective clinical worsening was not noted during follow-up.

CONCLUSION

With the increasing popularity of leadless pacemakers, it is paramount for device implanting teams to be familiar with common predictors of AV synchrony and troubleshooting with MicraTM -AV devices.

Для цитирования: Kowlgi Gurukripa N, Tseng Andrew S, Tempel Nathan D, Henrich Mark J, Venkatachalam Kalpathi L, Scott Luis, Shen Win-Kuang, Deshmukh Abhishek J, Madhavan Malini, Lee Hon-Chi, Asirvatham Samuel J, Friedman Paul A, Cha Yong-Mei, Mulpuru Siva K. A real-world experience of atrioventricular synchronous pacing with leadless ventricular pacemakers.. Journal of cardiovascular electrophysiology. 2022. DOI: 10.1111/jce.15430.