A real-world experience of atrioventricular synchronous pacing with leadless ventricular pacemakers.
Аннотация
AIMS: The MicraTM transcatheter pacing system (TPS) (Medtronic) is the only leadless pacemaker that promotes atrioventricular (AV) synchrony via accelerometer-based atrial sensing. Data regarding the real-world experience with this novel system are scarce. We sought to characterize patients undergoing MicraTM -AV implants, describe percentage AV synchrony achieved, and analyze the causes for suboptimal AV synchrony.
METHODS
In this retrospective cohort study, electronic medical records from 56 consecutive patients undergoing MicraTM -AV implants at the Mayo Clinic sites in Minnesota, Florida, and Arizona with a minimum follow-up of 3 months were reviewed. Demographic data, comorbidities, echocardiographic data, and clinical outcomes were compared among patients with and without atrial synchronous ventricular pacing (AsVP) ≥ 70%.
RESULTS
Sixty-five percent of patients achieved AsVP ≥ 70%. Patients with adequate AsVP had smaller body mass indices, a lower proportion of congestive heart failure, and prior cardiac surgery. Echocardiographic parameters and procedural characteristics were similar across the two groups. Active device troubleshooting was associated with higher AsVP. The likely reasons for low AsVP were small A4-wave amplitude, high ventricular pacing burden, and inadequate device reprogramming. Importantly, in patients with low AsVP, subjective clinical worsening was not noted during follow-up.
CONCLUSION
With the increasing popularity of leadless pacemakers, it is paramount for device implanting teams to be familiar with common predictors of AV synchrony and troubleshooting with MicraTM -AV devices.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.