DOI 10.1016/j.cardfail.2021.12.019
The V-LAP System for Remote Left Atrial Pressure Monitoring of Patients With Heart Failure: Remote Left Atrial Pressure Monitoring.
Аннотация
OBJECTIVE
Patients with heart failure (HF) are at an increased risk of hospital admissions. The aim of this report is to describe the feasibility, safety and accuracy of a novel wireless left atrial pressure (LAP) monitoring system in patients with HF.
METHODS
The V-LAP Left Atrium Monitoring systEm for Patients With Chronic sysTOlic & Diastolic Congestive heart Failure (VECTOR-HF) study is a prospective, multicenter, single-arm, open-label, first-in human clinical trial to assess the safety, performance and usability of the V-LAP system (Vectorious Medical Technologies) in patients with New York Heart Association class III HF. The device was implanted in the interatrial septum via a percutaneous, trans-septal approach guided by fluoroscopy and echocardiography. Primary endpoints included the successful deployment of the implant, the ability to perform initial pressure measurements and safety outcomes.
RESULTS
To date, 24 patients have received implants of the LAP-monitoring device. No device-related complications have occurred. LAP was reported accurately, agreeing well with wedge pressure at 3 months (Lin concordance correlation coefficient = 0.850). After 6 months, New York Heart Association class improved in 40% of the patients (95% CI = 16.4%-63.5%), while the 6-minute walk test distance had not changed significantly (313.9 ± 144.9 vs 232.5 ± 129.9 meters; P = 0.076).
CONCLUSION
The V-LAP left atrium monitoring system appears to be safe and accurate.