MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jacep.2020.07.021

Медицинская библиотека · JACC. Clinical electrophysiology · 2020

Clinical Presentation, Timing, and Microbiology of CIED Infections: An Analysis of the WRAP-IT Trial.

Sohail M Rizwan, Corey G Ralph, Wilkoff Bruce L, Poole Jeanne E, Mittal Suneet, Kennergren Charles, Greenspon Arnold J, Cheng Alan, Lande Jeffrey D, Lexcen Daniel R, Tarakji Khaldoun G

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study characterized the microbiology of major cardiac implantable electronic device (CIED) infections that occurred during the WRAP-IT (Worldwide Randomized Antibiotic Envelope Infection Prevention Trial) study.

BACKGROUND

The WRAP-IT study offers a unique opportunity for further understanding of the pathogens involved in major CIED infections in a prospective dataset, with implications for clinical practice and infection management.

METHODS

A total of 6,800 patients randomized 1:1 to receive an antibacterial envelope or not (control subjects) were included in this analysis. Patient characteristics, infection manifestation (pocket vs. systemic), and infection microbiology were evaluated through all follow-up (36 months). Data were analyzed using Cox proportional hazards regression.

RESULTS

A total of 3,371 patients received an envelope, and 3,429 patients were control subjects. Major CIED infection occurred in 32 patients who received an envelope and 51 control subjects (36-month Kaplan-Meier estimated event rate, 1.3% and 1.9%, respectively; p = 0.046). A 61% reduction in major pocket infection was observed within 12 months of the procedure in the envelope group (hazard ratio: 0.39, 95% confidence interval: 0.21 to 0.73; p = 0.003). Among 76 patients with major infections who had a sample taken, causative pathogens were identified in 47 patients. Staphylococcus species were the predominate pathogen (n = 31) and envelope use resulted in a 76% reduction in Staphylococcus-related pocket infections (n = 4 vs. 17; p = 0.010). Envelope use was not associated with delayed onset of pocket infections and did not affect the presentation of infections.

CONCLUSIONS

Antibacterial envelope use resulted in a significant reduction of major CIED pocket infections and was particularly effective against Staphylococcus species, the predominant cause of pocket infections. (Worldwide Randomized Antibiotic Envelope Infection Prevention Trial [WRAP-IT]; NCT02277990).

Для цитирования: Sohail M Rizwan, Corey G Ralph, Wilkoff Bruce L, Poole Jeanne E, Mittal Suneet, Kennergren Charles, Greenspon Arnold J, Cheng Alan, Lande Jeffrey D, Lexcen Daniel R, Tarakji Khaldoun G. Clinical Presentation, Timing, and Microbiology of CIED Infections: An Analysis of the WRAP-IT Trial.. JACC. Clinical electrophysiology. 2020. DOI: 10.1016/j.jacep.2020.07.021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.