DOI 10.1016/j.jstrokecerebrovasdis.2023.107430
A smartphone pupillometry tool for detection of acute large vessel occlusion.
Аннотация
OBJECTIVES
Pupillary light reflex (PLR) parameters can be used as quantitative biomarkers of neurological function. Since digital infrared pupillometry is expensive, we sought to examine alterations in PLR parameters using a smartphone quantitative pupillometry platform in subjects with acute ischemic stroke (AIS).
MATERIALS AND
METHODS
Patients were enrolled if they presented to the emergency department as a stroke code activation and had evidence of a large vessel occlusion (LVO) on computed tomography angiography. Controls were enrolled from hospital staff. A smartphone pupillometer was used in AIS patients with LVO pre-mechanical thrombectomy, immediately post-thrombectomy, and at 24 h post-thrombectomy. Clinical and demographic data were collected, along with the proprietary Neurological Pupil index (NPi) score from the NPi-200 digital infrared pupillometer. PLR parameters were compared using mean differences. The absolute and non-absolute inter-eye difference in each parameter for each subject were also analyzed by measuring 1 - (R:L) to determine alteration in the equilibrium between subject pupils. The NPi was analyzed for mean differences between cohorts.
RESULTS
Healthy controls (n = 132) and AIS patients (n = 31) were enrolled. Significant differences were observed in PLR parameters for healthy subjects when compared to pre-thrombectomy subjects in both mean and absolute inter-eye differences after post hoc Bonferroni correction. The proprietary NPi score was not significantly different for all groups and comparisons.
CONCLUSIONS
Significant alterations in the PLR were observed in AIS patients with LVO before thrombectomy, indicating the potential use of smartphone pupillometry for detection of LVO.