Hemodynamic Effects of Mechanical Circulatory Support Devices in Ventricular Septal Defect.

Pahuja Mohit, Schrage Benedikt, Westermann Dirk, Basir Mir B, Garan Arthur Reshad, Burkhoff Daniel

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Ventricular septal defect (VSD) is a lethal complication of acute myocardial infarction (AMI) and is often associated with cardiogenic shock. The optimal form of percutaneous mechanical circulatory support (MCS) for AMI-VSD is unknown.

METHODS AND

RESULTS

We used a previously validated cardiovascular model to simulate AMI-VSD with parameters adjusted to replicate average hemodynamics reported in the literature, including a pulmonary-to-systemic blood flow ratio of 3.0. We then predicted effects of different types of percutaneous MCS (including intra-aortic balloon pumping, Impella, TandemHeart, and extracorporeal membrane oxygenation) on pressures and flows throughout the cardiovascular system. The simulation replicated all major hemodynamic parameters reported in the literature with AMI-VSD. Inotropes and vasopressors worsened left-to-right shunting, whereas vasodilators decreased shunting at the expense of worsening hypotension. All MCS devices increased forward blood flow and arterial pressure but other effects varied among devices. Impella 5.0 provided the greatest degree of pulmonary capillary wedge pressure reductions and decreased left-to-right shunting. Extracorporeal membrane oxygenation worsened pulmonary capillary wedge pressure and shunting, which could be improved by adding Impella or passive left ventricular vent. Pulmonary-to-systemic blood flow ratio could not be reduced below 2.0, and pulmonary flows remained high with all forms of MCS.

CONCLUSIONS

Although no form of percutaneous MCS normalized hemodynamics in AMI-VSD, pulmonary capillary wedge pressure and shunting were worsened by extracorporeal membrane oxygenation and improved by Impella. Accordingly, based on hemodynamics alone, Impella provides the optimal form of support in AMI-VSD. However, other factors, including team experience, device availability, potential for tissue ingestion, and clinical characteristics, need to be considered when choosing a percutaneous MCS device for AMI-VSD.

Для цитирования: Pahuja Mohit, Schrage Benedikt, Westermann Dirk, Basir Mir B, Garan Arthur Reshad, Burkhoff Daniel. Hemodynamic Effects of Mechanical Circulatory Support Devices in Ventricular Septal Defect.. Circulation. Heart failure. 2019. DOI: 10.1161/circheartfailure.119.005981.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.