DOI 10.1007/s12194-025-00929-8

Медицинская библиотека · Radiological physics and technology · 2025

Clinical technique: evaluation of detection accuracy in surface-guided radiation therapy using CT references from different treatment planning systems.

Saito Masahide, Ueda Koji, Nemoto Hikaru, Tozuka Ryota, Miyasaka Yosuke, Onishi Yoshiko, Sugiyama Syuichiro, Sasada Yumi, Sano Naoki, Onishi Hiroshi

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

This study evaluated the detection accuracy of the VOXELAN surface-guided radiation therapy (SGRT) system using CT reference body surfaces generated by different radiation treatment planning systems (RTPSs) under their respective default CT value threshold settings. Two phantoms were used to assess 6-axis position matching accuracy with 1 mm and 2 mm slice thicknesses. Contour variations of approximately 2 mm were observed in the ventral and dorsal directions due to differing RTPSs. VOXELAN generally achieved detection accuracy within 1 mm, though some RTPS combinations showed errors exceeding 1 mm in the longitudinal and lateral directions. Slice thickness differences had minimal impact on detection accuracy. Overall, VOXELAN's detection accuracy varied slightly depending on the RTPS used but remained within approximately 1 mm. From our results, using a consistent RTPS when performing SGRT is recommended, as detection errors associated with different RTPS combinations were complex and difficult to interpret.

Для цитирования: Saito Masahide, Ueda Koji, Nemoto Hikaru, Tozuka Ryota, Miyasaka Yosuke, Onishi Yoshiko, Sugiyama Syuichiro, Sasada Yumi, Sano Naoki, Onishi Hiroshi. Clinical technique: evaluation of detection accuracy in surface-guided radiation therapy using CT references from different treatment planning systems.. Radiological physics and technology. 2025. DOI: 10.1007/s12194-025-00929-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.