MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.gie.2007.03.1049

Медицинская библиотека · Gastrointestinal endoscopy · 2007

Comparative study of endoscopic clips: duration of attachment at the site of clip application.

Shin Eun Ji, Ko Chung-Wang, Magno Priscilla, Giday Samuel A, Clarke John O, Buscaglia Jonathan M, Sedrakyan Gevorg, Jagannath Sanjay B, Kalloo Anthony N, Kantsevoy Sergey V

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Several designs of endoscopic clips are now commercially available, and the indications for endoclip application are rapidly expanding. However, very limited data have been published to aid in choosing between the different types of endoclips.

OBJECTIVE

To compare the duration of clip attachment between all commercially available endoclips.

SETTING

Long-term experiments on 50-kg pigs under general anesthesia.

DESIGN AND INTERVENTIONS: Upper endoscope was inserted into the stomach. One clip of each type (Resolution clip, TriClip, and HX-5L clip) was placed along the same gastric fold at a distance of 0.5 to 1 cm from each other. The animals were recovered. In pig nos. 1 and 2, repeat endoscopy was performed after 2 and 4 weeks. In pig nos. 3 to 5, endoscopy was repeated after 1, 2, and 5 weeks.

MAIN OUTCOME MEASUREMENTS: Duration of clip retention at the site of application.

RESULTS

In all animals, only the Resolution endoclip remained attached to the site of application for the entire duration of the study (4-5 weeks). No TriClips or HX-5L clips were attached at the 4- to 5-week follow-up endoscopies. Most of the TriClips (67%) detached within the first week after application. Most of the HX-5L clips (80%) dislodged within the first 2 weeks of follow-up.

LIMITATIONS: The study was performed in a porcine model with a small number of animals.

CONCLUSIONS

The Resolution clip has the longest duration of retention at the site of application (more than 4-5 weeks) and should be preferred when long-term attachment of endoclips is necessary.

Для цитирования: Shin Eun Ji, Ko Chung-Wang, Magno Priscilla, Giday Samuel A, Clarke John O, Buscaglia Jonathan M, Sedrakyan Gevorg, Jagannath Sanjay B, Kalloo Anthony N, Kantsevoy Sergey V. Comparative study of endoscopic clips: duration of attachment at the site of clip application.. Gastrointestinal endoscopy. 2007. DOI: 10.1016/j.gie.2007.03.1049.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.