HbA1c screening in the community: Lessons for safety and quality management of a point of care programme.

Kenealy Timothy, Herd Geoffrey, Musaad Samarina, Wells Susan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIM: To describe quality management processes and appropriate interpretation with respect to HbA1c point-of-care (POC) testing in a national diabetes and cardiovascular risk screening programme.

METHODS

We compared HbA1c results from capillary blood, measured by the cobas b 101 (Roche Diagnostics) POC testing system, with results from venous blood measured by accredited laboratory analysers to inform national screening practice and a (separately-reported) randomised controlled trial. Difference plots and regressions were used to aid interpretation around 40 and 50mmol/mol, the cut-offs used to identify "pre-diabetes" and diabetes in New Zealand.

RESULTS

After initial acceptable tests, subsequent batches delivered POC results that varied from laboratory HbA1c by +6 to -14mmol/mol around the clinical cut-offs. Ten faulty batches of discs were recalled worldwide. POC testing was suspended in one region, as was the planned trial. The manufacturing defect was rectified, accuracy of the new batches was confirmed, and testing resumed.

CONCLUSION

POC testing must be conducted within stringent quality assurance processes prior to and while in use. Within such a system, POC testing for HbA1c can be sufficiently accurate for screening and diagnosis of diabetes.

Для цитирования: Kenealy Timothy, Herd Geoffrey, Musaad Samarina, Wells Susan. HbA1c screening in the community: Lessons for safety and quality management of a point of care programme.. Primary care diabetes. 2018. DOI: 10.1016/j.pcd.2018.11.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.