MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvir.2013.08.014

Медицинская библиотека · Journal of vascular and interventional radiology : JVIR · 2013

Irreversible electroporation of the pig kidney with involvement of the renal pelvis: technical aspects, clinical outcome, and three-dimensional CT rendering for assessment of the treatment zone.

Sommer Christof M, Fritz Stefan, Wachter Miguel F, Vollherbst Dominik, Stampfl Ulrike, Bellemann Nadine, Gockner Theresa, Mokry Theresa, Gnutzmann Daniel, Schmitz Anne, Knapp Jürgen, Longerich Thomas, Kuhn-Neureuther Clemens, Pereira Philippe L, Kauczor Hans U, Werner Jens, Radeleff Boris A

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To analyze irreversible electroporation (IRE) of the pig kidney with involvement of the renal pelvis.

MATERIALS AND

METHODS

IRE of renal tissue including the pelvis was performed in 10 kidneys in five pigs. Three study groups were defined: group I (two applicators with parallel configuration; n = 11), group II (three applicators with triangular configuration; n = 2), and group III (six applicators with complex configuration; n = 3). After IRE and before euthanasia, pigs underwent contrast-enhanced computed tomography (CT). Technical aspects (radial distance of applicators, resulting mean current), clinical outcome (complications, blood samples), and three-dimensional CT rendering for assessment of the treatment zone (short axis, circularity) were assessed.

RESULTS

Radial distances of applicators were 14.3 mm ± 2.8 in group I, 12.3 mm ± 1.9 in group II, and 16.4 mm ± 3.5 in group III. Resulting mean currents were 25.7 A ± 6.5 in group I, 27.0 A ± 7.1 in group II, and 39.4 A ± 8.9 in group III. In group III, two perirenal hematomas were identified. There was no damage to the renal pelvis. During IRE, clinical blood parameters and cardiovascular markers did not change significantly. Short axis measurements were 20.6 mm ± 3.6 in group I, 31.9 mm ± 8.2 in group II, and 39.3 mm ± 2.4 in group III (P < .01 between groups). Circularity scores were 0.8 ± 0.2 in group I, 0.7 ± 0.1 in group II, and 0.7 ± 0.1 in group III, with a score of 1 indicating perfect roundness (P value not significant).

CONCLUSIONS

IRE of the pig kidney with involvement of the renal pelvis is feasible and safe. Size but not shape of the treatment zone is significantly affected by applicator configuration.

Для цитирования: Sommer Christof M, Fritz Stefan, Wachter Miguel F, Vollherbst Dominik, Stampfl Ulrike, Bellemann Nadine, Gockner Theresa, Mokry Theresa, Gnutzmann Daniel, Schmitz Anne, Knapp Jürgen, Longerich Thomas, Kuhn-Neureuther Clemens, Pereira Philippe L, Kauczor Hans U, Werner Jens, Radeleff Boris A. Irreversible electroporation of the pig kidney with involvement of the renal pelvis: technical aspects, clinical outcome, and three-dimensional CT rendering for assessment of the treatment zone.. Journal of vascular and interventional radiology : JVIR. 2013. DOI: 10.1016/j.jvir.2013.08.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.