The Journal of urology · 2023
First in Human Subjects Testing of the UroMonitor: A Catheter-free Wireless Ambulatory Bladder Pressure Monitor.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Frainey Brendan T, Majerus Steve J A, Derisavifard Samir, Lewis Kevin C, Williams Anna R, Balog Brian M, Butler Robert S, Goldman Howard B, Damaser Margot S
Аннотация
PURPOSE: Urodynamics is the standard method of diagnosing bladder dysfunction, but involves catheters and retrograde bladder filling. With these artificial conditions, urodynamics cannot always reproduce patient complaints. We have developed a wireless, catheter-free intravesical pressure sensor, the UroMonitor, which enables catheter-free telemetric ambulatory bladder monitoring. The purpose of this study was twofold: to evaluate accuracy of UroMonitor pressure data, and assess safety and feasibility of use in humans.
MATERIALS AND METHODS: Eleven adult female patients undergoing urodynamics for overactive bladder symptoms were enrolled. After baseline urodynamics, the UroMonitor was transurethrally inserted into the bladder and position was confirmed cystoscopically. A second urodynamics was then performed with the UroMonitor simultaneously transmitting bladder pressure. Following removal of urodynamics catheters, the UroMonitor transmitted bladder pressure during ambulation and voiding in private. Visual analogue pain scales (0-5) were used to assess patient discomfort.
RESULTS: The UroMonitor did not significantly alter capacity, sensation, or flow during urodynamics. The UroMonitor was also easily inserted and removed in all subjects. The UroMonitor reproduced bladder pressure, capturing 98% (85/87) of voiding and nonvoiding urodynamic events. All subjects voided with only the UroMonitor in place with low post-void residual volume. Median ambulatory pain score with the UroMonitor was rated 0 (0-2). There were no post-procedural infections or changes to voiding behavior.
CONCLUSIONS: The UroMonitor is the first device to enable catheter-free telemetric ambulatory bladder pressure monitoring in humans. The UroMonitor appears safe and well tolerated, does not impede lower urinary tract function, and can reliably identify bladder events compared to urodynamics.
MATERIALS AND METHODS: Eleven adult female patients undergoing urodynamics for overactive bladder symptoms were enrolled. After baseline urodynamics, the UroMonitor was transurethrally inserted into the bladder and position was confirmed cystoscopically. A second urodynamics was then performed with the UroMonitor simultaneously transmitting bladder pressure. Following removal of urodynamics catheters, the UroMonitor transmitted bladder pressure during ambulation and voiding in private. Visual analogue pain scales (0-5) were used to assess patient discomfort.
RESULTS: The UroMonitor did not significantly alter capacity, sensation, or flow during urodynamics. The UroMonitor was also easily inserted and removed in all subjects. The UroMonitor reproduced bladder pressure, capturing 98% (85/87) of voiding and nonvoiding urodynamic events. All subjects voided with only the UroMonitor in place with low post-void residual volume. Median ambulatory pain score with the UroMonitor was rated 0 (0-2). There were no post-procedural infections or changes to voiding behavior.
CONCLUSIONS: The UroMonitor is the first device to enable catheter-free telemetric ambulatory bladder pressure monitoring in humans. The UroMonitor appears safe and well tolerated, does not impede lower urinary tract function, and can reliably identify bladder events compared to urodynamics.