MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.15132

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2021

Safety and efficacy of left atrial appendage closure using an epicardial suture snaring device: Systematic review and current status.

Sabzwari Syed Rafay Ali, Mehta Nikhil A, Garg Lohit, Racharla Lekha, Kurtz Emilee, Afzal Muhammad R, Turagam Mohit K, Lakkireddy Dhanunjaya

СтатьяСистематический обзорПеревод пока не подготовлен

Аннотация

INTRODUCTION

The LARIAT epicardial suture snaring device has been fraught with technical challenges and procedural complications. Over time, technique modification and improved operator experience have helped overcome many of these challenges. METHODS AND

RESULTS

Studies reporting left atrial appendage epicardial closure over a 12-year period from 2009 to 2020 were evaluated. The primary efficacy outcome evaluated was left atrial appendage closure with residual flow ≤1 mm. Safety outcomes evaluated were periprocedural and long term complications after device placement. Nine studies consisting of 1430 patients were included in this analysis, of which 1386 (97%) underwent successful device placement. The mean age was 69.3 years, with a mean CHADS2 -VASC score of 3.78. Primary efficacy end point was achieved in 95.8% patients immediately after device placement, and in 92.8% patients on long term follow-up. Early procedural complications consisted of pericardial effusion 5.6%, cardiac perforation 2.7%, pericarditis 2.6%, and need for open heart surgery 1.5%. Long term complications consisted of all-cause mortality 2.3%, stroke 1.5%, and left atrial appendage thrombus 2.3%. Reduced complications were noted using micropuncture needles (2.20% vs. 10.14%; p < .0001), a longer duration pericardial drainage, and use of anti-inflammatory medications (1.58% vs. 8.4%). Oral anticoagulation use decreased from 44.7% to 22.9% post device implantation, and to 8.5% on last clinical follow up.

CONCLUSIONS

The LARIAT device is effective in epicardial closure of the left atrial appendage. Improvement in device techniques such as use of micropuncture needle, prophylactic colchicine, and maintenance of a pericardial drain have helped improve safety over time.

Для цитирования: Sabzwari Syed Rafay Ali, Mehta Nikhil A, Garg Lohit, Racharla Lekha, Kurtz Emilee, Afzal Muhammad R, Turagam Mohit K, Lakkireddy Dhanunjaya. Safety and efficacy of left atrial appendage closure using an epicardial suture snaring device: Systematic review and current status.. Journal of cardiovascular electrophysiology. 2021. DOI: 10.1111/jce.15132.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.