Neosinus and Sinus Flow After Self-Expanding and Balloon-Expandable Transcatheter Aortic Valve Replacement.

Hatoum Hoda, Gooden Shelley C M, Sathananthan Janarthanan, Sellers Stephanie, Kutting Maximilian, Marx Philipp, Lilly Scott M, Ihdayhid Abdul R, Thourani Vinod H, Dasi Lakshmi Prasad

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The aim of this study was to evaluate flow dynamics in the aortic sinus and the neosinus (NS) after transcatheter heart valve (THV) implantation in valve-in-valve (ViV).

BACKGROUND

Leaflet thrombosis may occur on THVs and affect performance and durability. Differences in flow dynamics may affect the risk for leaflet thrombosis.

METHODS

Hemodynamic assessment following THV implantation in a surgical aortic valve was performed in a left heart simulator under pulsatile physiological conditions. Assessment was performed using a 23-mm polymeric surgical aortic valve (not diseased) and multiple THV platforms, including self-expanding devices (26-mm Evolut, 23-mm Allegra, small ACURATE neo) and a balloon-expandable device (23-mm SAPIEN 3). Particle image velocimetry was performed to assess flow in the sinus and NS. Sinus and NS washout, shear stress, and velocity were calculated.

RESULTS

Sinus and NS washout was fastest and approximately 1 cardiac cycle for each with the Evolut, ACURATE neo, and Allegra compared with the SAPIEN 3, with washout in 2 and 3 cardiac cycles, respectively. The Allegra showed the largest shear stress distribution in the sinus, followed by the SAPIEN 3. In the NS, all 4 valves showed equal likelihoods of occurrence of shear stress <1 Pa, but the Allegra showed the highest likelihoods of occurrence for shear stress >1 Pa. The velocities in the sinus and NS were 0.05, 0.078, 0.080, and 0.075 m/s for Evolut, SAPIEN 3, ACURATE neo, and Allegra ViV, respectively.

CONCLUSIONS

Sinus and NS flow dynamics differ substantially among THVs after ViV. Self-expanding supra-annular valves seem to have faster washouts compared with an equivalent-size balloon-expandable THV.

Для цитирования: Hatoum Hoda, Gooden Shelley C M, Sathananthan Janarthanan, Sellers Stephanie, Kutting Maximilian, Marx Philipp, Lilly Scott M, Ihdayhid Abdul R, Thourani Vinod H, Dasi Lakshmi Prasad. Neosinus and Sinus Flow After Self-Expanding and Balloon-Expandable Transcatheter Aortic Valve Replacement.. JACC. Cardiovascular interventions. 2021. DOI: 10.1016/j.jcin.2021.09.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.