Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead?

van Everdingen Wouter M, Zweerink Alwin, Cramer Maarten J, Doevendans Pieter A, Nguyên Uyên Châu, van Rossum Albert C, Prinzen Frits W, Vernooy Kevin, Allaart Cornelis P, Meine Mathias

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Previous studies indicated the importance of the intrinsic left ventricular (LV) electric delay (QLV) for optimal benefit to cardiac resynchronization therapy. We investigated the use of QLV for achieving optimal acute hemodynamic response to cardiac resynchronization therapy with a quadripolar LV lead.

METHODS AND

RESULTS

Forty-eight heart failure patients with a left bundle branch block were prospectively enrolled (31 men; age, 66±10 years; LV ejection fraction, 28±8%; QRS duration, 176±14 ms). Immediately after cardiac resynchronization therapy implantation, invasive LV pressure-volume loops were recorded during biventricular pacing with each separate electrode at 4 atrioventricular delays. Acute cardiac resynchronization therapy response, measured as change in stroke work (Δ%SW) compared with intrinsic conduction, was related to intrinsic interval between Q on the ECG and LV sensing delay (QLV), normalized for QRS duration (QLV/QRSd), and electrode position. QLV/QRSd was 84±9% and variation between the 4 electrodes 9±5%. Δ%SW was 89±64% and varied by 39±36% between the electrodes. In univariate analysis, an anterolateral or lateral electrode position and a high QLV/QRSd had a significant association with a large Δ%SW (all P <0.01). In a combined model, only QLV/QRSd remained significantly associated with Δ%SW (P<0.05). However, a direct relation between QLV/QRSd and Δ%SW was only seen in 24 patients, whereas 24 patients showed an inverse relation.

CONCLUSIONS

The large variation in acute hemodynamic response indicates that the choice of the stimulated electrode on a quadripolar lead is important. Although QLV/QRSd was associated with acute hemodynamic response at group level, it cannot be used to select the optimal electrode in the individual patient.

Для цитирования: van Everdingen Wouter M, Zweerink Alwin, Cramer Maarten J, Doevendans Pieter A, Nguyên Uyên Châu, van Rossum Albert C, Prinzen Frits W, Vernooy Kevin, Allaart Cornelis P, Meine Mathias. Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead?. Circulation. Arrhythmia and electrophysiology. 2018. DOI: 10.1161/circep.117.005912.