DOI 10.1016/j.hrthm.2018.05.027

Медицинская библиотека · Heart rhythm · 2018

Scar burden, not intraventricular conduction delay pattern, is associated with outcomes in ischemic cardiomyopathy patients receiving cardiac resynchronization therapy.

Adelstein Evan C, Althouse Andrew D, Schwartzman David, Jain Sandeep K, Soman Prem, Saba Samir

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Patients with nonspecific intraventricular conduction delay (IVCD) benefit less from cardiac resynchronization therapy (CRT) than patients with left bundle branch block (LBBB).

OBJECTIVE

The purpose of this study was to determine whether post-CRT outcome differences in patients with ischemic cardiomyopathy (ICM) relate to intrinsic QRS pattern and/or scar burden.

METHODS

We analyzed 393 consecutive ICM patients with left ventricular ejection fraction (LVEF) ≤35%, QRS duration >120 ms, and LBBB or nonspecific IVCD who underwent single-photon emission computed tomography myocardial perfusion imaging and CRT-defibrillator implant. We compared scar burden; QRS duration; LVEF change; risk of death, transplant, or ventricular assist device; and risk of appropriate device shocks between LBBB and IVCD patients, using multivariable analyses to determine relative associations between QRS pattern vs scar burden and outcomes.

RESULTS

Nonspecific IVCD is associated with greater scar burden and narrower baseline QRS duration than LBBB. IVCD patients demonstrated less QRS narrowing with CRT than LBBB patients, even when excluding IVCD patients with QRS duration <150 ms. LVEF improved less in patients with IVCD vs LBBB, but only scar burden not QRS morphology or duration was associated with LVEF increase ≥5%. During 39-month follow-up, IVCD was associated with shorter survival free from transplant/ventricular assist device and shorter time to first appropriate device shock. Scar burden but not QRS morphology was independently associated with these outcomes on multivariate analysis.

CONCLUSION

IVCD is associated with greater scar burden than LBBB in ICM CRT-defibrillator recipients. Scar burden, not QRS pattern, is independently associated with adverse clinical outcomes.

Для цитирования: Adelstein Evan C, Althouse Andrew D, Schwartzman David, Jain Sandeep K, Soman Prem, Saba Samir. Scar burden, not intraventricular conduction delay pattern, is associated with outcomes in ischemic cardiomyopathy patients receiving cardiac resynchronization therapy.. Heart rhythm. 2018. DOI: 10.1016/j.hrthm.2018.05.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.