DOI 10.1111/pace.13344

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2018

Feasibility of cathodic-anodal left ventricular stimulation for alternative multisite pacing.

Dell'Era Gabriele, De Vecchi Federica, Prenna Eleonora, Devecchi Chiara, Degiovanni Anna, Malacrida Maurizio, Magnani Andrea, Occhetta Eraldo, Marino Paolo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Simultaneous cathodic-anodal capture by a bipole of a cardiac resynchronization therapy (CRT) left-ventricular (LV) catheter may depolarize a larger LV area than conventional multipoint pacing. We evaluated the feasibility of cathodic-anodal LV stimulation.

METHODS

In 30 patients undergoing CRT with a quadripolar LV lead, we evaluated the cathodic and anodal capture threshold for each LV pole and compared QRS on electrocardiogram (ECG) during single-point cathodic biventricular stimulation (S-BS), multipoint BS (M-BS), and cathodic-anodal BS (CA-BS).

RESULTS

Anodal capture was obtained by three poles in 23/30 patients, by two poles in five, and was not feasible in two. The mean single-point anodal threshold was 3.93 V versus single-point cathodic threshold of 1.95 V. On comparing ECGs, M-BS and CA-BS produced similar QRS wavefront activation in 90% of patients.

CONCLUSIONS

CA-BS is feasible and may be used in LV pacing to achieve a different wavefront of electrical activation. Further prospective studies are needed in order to verify the clinical impact of this kind of stimulation.

Для цитирования: Dell'Era Gabriele, De Vecchi Federica, Prenna Eleonora, Devecchi Chiara, Degiovanni Anna, Malacrida Maurizio, Magnani Andrea, Occhetta Eraldo, Marino Paolo. Feasibility of cathodic-anodal left ventricular stimulation for alternative multisite pacing.. Pacing and clinical electrophysiology : PACE. 2018. DOI: 10.1111/pace.13344.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.