'Optimized' LV only pacing using a dual chamber pacemaker as a cost effective alternative to CRT.

Rai Maneesh K, Prabhu Mukund A, Sharma Abhishek, Vekariya Ritesh, Kamath Padmanabh, Pai Narasimha, Kamath Ramanath L

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiac Resynchronization therapy (CRT) remains largely under-used in developing countries owing to the high cost of therapy. In this pilot study, we explore 'optimized' Left Ventricle Only Pacing (LVOP) as a cost effective alternative to cardiac resynchronization therapy in selected patients with heart failure.

HYPOTHESIS: In economically poorer patients with heart failure, left bundle branch block (LBBB) and intact AV node conduction, synchronization can be obtained using a dual chamber pacemaker (leads in right atrium and Left ventricle) with the help of 2D strain imaging.

METHODS AND

RESULTS

4 patients underwent LVOP for symptomatic heart failure. Post procedure 'optimization' was done using 12 lead electrocardiography and 2D- Strain imaging. Difference between Time to Peak longitudinal strain and Aortic valve Closure (Diff TPL-AC) was calculated for each segment at different AV delays and the AV delay with the smallest Diff TPL-AC was programmed. The mean AV delay that resulted in electrical and mechanical synchrony was 150 ms. After a mean follow up of 6 months, all patients had improved by at least 1 NYHA class. The mean reduction in QRS duration post procedure was -54.5 ± 22.82 ms and the mean improvement in EF was 7 ± 2.75%.

CONCLUSION

Optimized LVOP using 2D strain and ECG can be a cost-effective alternative to CRT in patients with LBBB, heart failure and normal AV node conduction.

Для цитирования: Rai Maneesh K, Prabhu Mukund A, Sharma Abhishek, Vekariya Ritesh, Kamath Padmanabh, Pai Narasimha, Kamath Ramanath L. 'Optimized' LV only pacing using a dual chamber pacemaker as a cost effective alternative to CRT.. Indian pacing and electrophysiology journal. 2017. DOI: 10.1016/j.ipej.2017.05.001.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.