Maximization of interventricular conduction time by means of quadripolar leads for cardiac resynchronization therapy.

Stabile Giuseppe, Bianchi Valter, Solimene Francesco, Iuliano Assunta, Parisi Quintino, Pepi Patrizia, Bocchiardo Mario, Urraro Francesco, De Simone Antonio, Ospizio Roberto, D'Onofrio Antonio

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Identifying the left ventricular (LV) site associated with the maximum spontaneous interventricular conduction time (right ventricle (RV)-to-LV interval) has proved to be an effective strategy for optimal LV pacing site selection in cardiac resynchronization therapy (CRT). The aim of our study was to determine whether quadripolar LV lead technology allows RV-to-LV interval maximization.

METHODS

We enrolled 108 patients undergoing implantation of a CRT system using an LV quadripolar lead and 114 patients who received a bipolar lead. On implantation, the RV-to-LV interval was measured for the dipole of the bipolar leads and for each electrode of the LV lead (tip, ring 2, ring 3, ring 4).

RESULTS

In the quadripolar group, the mean RV-to-LV interval ranged from 90 ± 33 ms (tip) to 94 ± 32 ms (R4) (p > 0.05 for all comparisons). In 55 (51%) patients, the RV-to-LV interval was > 80 ms at all electrodes, while in 27 (25%) patients, no electrodes were associated with an RV-to-LV interval > 80 ms. At least one LV pacing electrode was associated with an RV-to-LV interval > 80 ms in 62 (70%) patients with a short (36 mm) inter-electrode distance, and in 19 (95%, p = 0.022) of those with a long distance (50.5 mm). In the bipolar group, the mean RV-to-LV interval was 72 ± 37 ms (p < 0.001 versus quadripolar). The RV-to-LV interval was > 80 ms in 44 (39%) patients (p < 0.001 versus quadripolar leads with both short and long inter-electrode distance).

CONCLUSIONS

Quadripolar leads allow RV-to-LV interval maximization. An optimal RV-to-LV interval seems achievable in the majority of patients, especially if the leads present a long inter-electrode distance.

Для цитирования: Stabile Giuseppe, Bianchi Valter, Solimene Francesco, Iuliano Assunta, Parisi Quintino, Pepi Patrizia, Bocchiardo Mario, Urraro Francesco, De Simone Antonio, Ospizio Roberto, D'Onofrio Antonio. Maximization of interventricular conduction time by means of quadripolar leads for cardiac resynchronization therapy.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2017. DOI: 10.1007/s10840-017-0279-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.