Novel fat suppression technique for ultrashort echo time MRI using single-point Dixon phase modeling.
Аннотация
BACKGROUND
Fat suppression plays a vital role in numerous magnetic resonance imaging (MRI) examinations, particularly in the musculoskeletal (MSK) system. However, current fat suppression methods are not fully optimized for ultrashort echo time (UTE) imaging, despite being essential for many advanced UTE-based imaging applications. This study aimed to investigate a novel fat suppression technique for UTE MRI using a single-point Dixon (1p-Dixon) approach through phase modeling.
METHODS
In this study, four cadaveric human knee joints, and six healthy volunteers were included. A 1p-Dixon-based fat suppression method was developed, which utilizes intrinsic information from complex UTE signals. Additionally, a data-driven approach based on the phase distribution was used for the decomposition of water and fat signals in short T2 tissues. The feasibility of the proposed method was evaluated in a fat-water phantom first and validated in ex vivo and in vivo human knee joints. The patella tendon, cartilage, posterior cruciate ligament (PCL), anterior cruciate ligament (ACL), and meniscus were evaluated in each knee.
RESULTS
In the phantom experiment, there was a significant correlation between the estimated fat fraction and the actual fat fraction (R>0.98; P<0.05). The ex vivo experiment revealed a significant difference in contrast-to-noise ratios (CNRs) measured from the two images without and with 1p-Dixon (P<0.001). The CNR values ranged from 3.4±0.5 to 9.6±5.0 and 1.8±1.6 to 4.1±0.8 for measurement with and without 1p-Dixon, respectively. The 1p-Dixon significantly improved the contrast in the in vivo experiment (P<0.0001). The CNR values ranged from 5.1±6.0 to 41.0±9.7 and 2.7±1.2 to 15.4±3.3 for measurement with and without 1p-Dixon, respectively in the in vivo experiment.
CONCLUSIONS
Our novel fat suppression technique has been shown to provide a fast, time-saving, and robust fat suppression for UTE imaging without the need for additional scans.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.