Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study.

Inaba Shinji, Mintz Gary S, Yun Kyeong Ho, Yakushiji Tadayuki, Shimizu Takehisa, Kang Soo-Jin, Généreux Philippe, Weisz Giora, Rabbani Leroy E, Moses Jeffrey W, Stone Gregg W, Maehara Akiko

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Mechanical complications contribute to bare metal and first-generation drug-eluting stent (DES) failure. However, the importance of the mechanical complications of second-generation DES remains unclear. We report mechanical complications associated with everolimus-eluting stent (EES) failures.

METHODS AND

RESULTS

We retrospectively analysed 177 consecutive EES-treated lesions in 136 patients who underwent intravascular ultrasound (IVUS) at follow-up. Mechanical complications were identified in 17 patients (five stable angina, 10 unstable angina, two non-ST-elevation myocardial infarction [NSTEMI] without angiographic thrombus). Fifteen (88.2%) were treated with repeat revascularisation. By IVUS, there were 16 focal (94.1%) and one diffuse (5.9%) in-stent restenoses. Complete stent fracture with separation was seen in only one, partial stent fracture with separation was seen in three, and in 13 there was longitudinal deformation (n=2) or stent strut fracture (n=11) with overlapping of the proximal and distal stent fragments. In 13 EES with evidence of overlapping in the setting of either fracture or deformation, there was a 35.5±12.2% smaller stent area compared to the adjacent proximal and distal stent fragments, and >50% neointimal hyperplasia in 12 (92.3%).

CONCLUSIONS

We found EES mechanical complications, often followed by longitudinal deformation or fracture leading to excessive neointimal hyperplasia, in-stent restenosis, and repeat revascularisation.

Для цитирования: Inaba Shinji, Mintz Gary S, Yun Kyeong Ho, Yakushiji Tadayuki, Shimizu Takehisa, Kang Soo-Jin, Généreux Philippe, Weisz Giora, Rabbani Leroy E, Moses Jeffrey W, Stone Gregg W, Maehara Akiko. Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2014. DOI: 10.4244/eijv9i11a220.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.