JACC. Cardiovascular interventions · 2013

First-in-human evaluation of a bioabsorbable polymer-coated sirolimus-eluting stent: imaging and clinical results of the DESSOLVE I Trial (DES with sirolimus and a bioabsorbable polymer for the treatment of patients with de novo lesion in the native coronary arteries).

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ormiston John, Webster Mark, Stewart James, Vrolix Mathias, Whitbourn Robert, Donohoe Dennis, Knape Charlene, Lansky Alexandra, Attizzani Guilherme F, Fitzgerald Peter, Kandzari David E, Wijns William

Аннотация

OBJECTIVES: This first-in-human multicenter study sought to examine prospectively the safety and efficacy of a new, cobalt chromium thin-strut, coronary absorbable polymer-coated, sirolimus-eluting stent.

BACKGROUND: Bioabsorbable polymers on drug-eluting stents may lower the long-term risks of inflammation, delayed healing, and adverse events.

METHODS: We enrolled patients with symptomatic coronary artery disease with stable or unstable angina pectoris and >50% diameter stenosis, amenable to coverage with a ≤23-mm long stent in a vessel 2.5 to 3.5 mm in diameter. All patients received dual antiplatelet therapy after implantation. Patients, in groups of 10, underwent repeat angiography, intravascular ultrasound, and optical coherence tomography at 4, 6, or 8 months, and all patients were seen or contacted at 18 months of follow-up.

RESULTS: The median (range) in-stent late lumen loss (LLL) was 0.03 mm (-0.22 to 0.21 mm), 0.10 mm (-0.03 to 1.2 mm), and 0.08 mm (-0.01 to 0.28 mm), at 4, 6, and 8 months, respectively. At 18 months, the median in-stent LLL was 0.08 mm (-0.30 to 0.46 mm). On optical coherence tomography, the proportion of uncovered stent struts decreased from a median of 7.3% (range 0.4% to 46.3%) at 4 months to 0% (range: 0% to 3.4%) at 18 months. The percentage of neointimal volume obstruction by intravascular ultrasound increased from a median of 5.3% to 9.1% between 4 and 6 months and remained nearly unchanged thereafter through 18 months of follow-up. The only recorded major adverse cardiac event was a myocardial infarction.

CONCLUSIONS: At 18 months of follow-up, this absorbable polymer-coated, cobalt chromium sirolimus-eluting stent was associated with a low and stable in-stent LLL, complete strut coverage, and no stent thrombosis. (First-In-Human Trial of the MiStent Drug-Eluting Stent [DES] in Coronary Artery Disease [DESSOLVE-I]; NCT01247428).