Measurement of aldosterone in human plasma by semiautomated HPLC-tandem mass spectrometry.
СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Taylor Paul J, Cooper Donald P, Gordon Richard D, Stowasser Michael
Аннотация
BACKGROUND: Reliable measurement of aldosterone with less interlaboratory variation than RIA would help standardize testing for primary aldosteronism. We set out to validate a high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) method for aldosterone in human plasma.
METHODS: We prepared samples (EDTA plasma, lithium heparin plasma, and serum from separator and plain clot tubes) and measured aldosterone using online HPLC-MS/MS with d(7)-aldosterone as internal standard. We also analyzed EDTA plasma samples by immunoassay. We established a reference range for HPLC-MS/MS aldosterone by analyzing blood collected midmorning from 97 normotensive seated subjects.
RESULTS: The linear range was 69.4-5548.0 pmol/L (2.5-200 ng/dL) (r(2) > 0.994, n = 14). Inter- and intraday analytical recovery and imprecision for quality control samples of 166.4, 1109.6, and 4161.0 pmol/L (6.0, 40.0, and 150.0 ng/dL) were 92.2%-102.0% and <6.3%, respectively (n = 5). The lower limit of quantification was 69.4 pmol/L (2.5 ng/dL), with inter- and intraday analytical recovery and imprecision of 91.4%-94.5% and <9.5% (n = 5). No interferences were observed in plasma from Addison's disease patients (n = 5). Comparison of collection tubes, using EDTA as the reference, revealed similar aldosterone results. Comparison of HPLC-MS/MS with immunoassay gave an acceptable mean bias (0.83%) but wide range (-44.8% to 39.7%) of differences. HPLC-MS/MS aldosterone concentrations in normotensive subjects ranged from <69.4 to 635.2 pmol/L (<2.5 to 22.9 ng/dL).
CONCLUSIONS: This first reported aldosterone method using online HPLC-MS/MS is precise across the clinically relevant range, not influenced by collection tube type, and offers semiautomated sample preparation and high throughput.
METHODS: We prepared samples (EDTA plasma, lithium heparin plasma, and serum from separator and plain clot tubes) and measured aldosterone using online HPLC-MS/MS with d(7)-aldosterone as internal standard. We also analyzed EDTA plasma samples by immunoassay. We established a reference range for HPLC-MS/MS aldosterone by analyzing blood collected midmorning from 97 normotensive seated subjects.
RESULTS: The linear range was 69.4-5548.0 pmol/L (2.5-200 ng/dL) (r(2) > 0.994, n = 14). Inter- and intraday analytical recovery and imprecision for quality control samples of 166.4, 1109.6, and 4161.0 pmol/L (6.0, 40.0, and 150.0 ng/dL) were 92.2%-102.0% and <6.3%, respectively (n = 5). The lower limit of quantification was 69.4 pmol/L (2.5 ng/dL), with inter- and intraday analytical recovery and imprecision of 91.4%-94.5% and <9.5% (n = 5). No interferences were observed in plasma from Addison's disease patients (n = 5). Comparison of collection tubes, using EDTA as the reference, revealed similar aldosterone results. Comparison of HPLC-MS/MS with immunoassay gave an acceptable mean bias (0.83%) but wide range (-44.8% to 39.7%) of differences. HPLC-MS/MS aldosterone concentrations in normotensive subjects ranged from <69.4 to 635.2 pmol/L (<2.5 to 22.9 ng/dL).
CONCLUSIONS: This first reported aldosterone method using online HPLC-MS/MS is precise across the clinically relevant range, not influenced by collection tube type, and offers semiautomated sample preparation and high throughput.