DOI 10.1109/tbme.2022.3200626

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2023

Single-Sided Magnet System for Quantitative MR Relaxometry and Preclinical In-Vivo Monitoring.

Thomas Dion G, Tzeng Yu-Chieh, Galvosas Petrik, Harrison Freya G, Berry Mary J, Teal Paul D, Galvin Sean D, Obruchkov Sergei I

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

We have developed a single-sided magnet system that allows Magnetic Resonance relaxation and diffusion parameters to be measured.

METHODS

A single-sided magnet system has been developed, using an array of permanent magnets. The magnet positions are optimised to produce a B0 magnetic field with a spot that is relatively homogenous and can project into a sample. NMR relaxometry experiments are used to measure quantitative parameters such as T2, T1 and apparent diffusion coefficient (ADC) on samples on the benchtop. To explore preclinical application, we test whether it can detect changes during acute global cerebral hypoxia in an ovine model.

RESULTS

The magnet produces a 0.2 T field projected into the sample. Measurements of benchtop samples show that it can measure T1, T2 and ADC, producing trends and values that are in line with literature measurements. In-vivo studies show a decrease in T2 during cerebral hypoxia that recovers following normoxia.

CONCLUSION

The single-sided MR system has the potential to allow non-invasive measurements of the brain. We also demonstrate that it can operate in a pre-clinical environment, allowing T2 to be monitored during brain tissue hypoxia.

SIGNIFICANCE: MRI is a powerful technique for non-invasive diagnosis in the brain, but its application has been limited by the requirements for magnetic field strength and homogeneity that imaging methods have. The technology described in this study provides a portable alternative to acquiring clinically significant MR parameters without the need for traditional imaging equipment.

Для цитирования: Thomas Dion G, Tzeng Yu-Chieh, Galvosas Petrik, Harrison Freya G, Berry Mary J, Teal Paul D, Galvin Sean D, Obruchkov Sergei I. Single-Sided Magnet System for Quantitative MR Relaxometry and Preclinical In-Vivo Monitoring.. IEEE transactions on bio-medical engineering. 2023. DOI: 10.1109/tbme.2022.3200626.