DOI 10.1111/pace.14743

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2023

Unusual atrial pacing induced by type II far-field p-wave sensing in a patient with complete atrioventricular block.

Yamagishi Yasunobu, Oginosawa Yasushi, Ishii Nozomu, Nakamura Yuki, Yagyu Keishiro, Miyamoto Taro, Ohe Hisaharu, Kohno Ritsuko, Kataoka Masaharu, Abe Haruhiko

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

An 88-year-old Japanese woman underwent DDD pacemaker (MicroPort KORA 250 DR, V lead: VEGA R52) implantation for complete atrioventricular block. A 12-lead electrocardiogram for a routine examination showed atrial pacing within the intrinsic P wave, followed by inhibition of ventricular pacing. Pacemaker interrogation revealed no abnormalities in the basic parameters; however, ventricular pacing was inhibited by far-field sensing of intrinsic atrial waves before atrial events; type II far-field P-wave sensing. As a result, unusual atrial pacing occurred due to the pause suppression algorithm, which is the one of the functions that prevent atrial fibrillation development.

Для цитирования: Yamagishi Yasunobu, Oginosawa Yasushi, Ishii Nozomu, Nakamura Yuki, Yagyu Keishiro, Miyamoto Taro, Ohe Hisaharu, Kohno Ritsuko, Kataoka Masaharu, Abe Haruhiko. Unusual atrial pacing induced by type II far-field p-wave sensing in a patient with complete atrioventricular block.. Pacing and clinical electrophysiology : PACE. 2023. DOI: 10.1111/pace.14743.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.