Transcatheter aortic valve implantation of a second-generation valve for pure aortic regurgitation: procedural outcome, haemodynamic data and follow-up.

Schlingloff Friederike, Schäfer Ulrich, Frerker Christian, Schmoeckel Michael, Bader Ralf

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The second-generation Jenavalve prosthesis (Jenavalve Technology, Inc., Munich, Germany) is the first transcatheter valve Conformité Européene (CE) marked for treatment of both aortic stenosis (AS) and pure aortic regurgitation (AR). Although the feasibility of the Jenavalve transcatheter aortic valve implantation (TAVI) in patients with pure AR has been described, haemodynamic and follow-up data are lacking.

METHODS

We report on a series of 10 transapical Jenavalve implantations for pure AR between December 2012 and September 2013. The patients were determined for TAVI by heart team decision at high surgical risk [log EuroSCORE (European System for Cardiac Operative Risk Evaluation) >20%], frailty or Charlson Comorbidity Index (CCI). Transaortic gradients and right heart haemodynamics were measured invasively before and after TAVI. Ventriculography and transoesophageal echocardiography were used to determine paravalvular regurgitation. All-cause mortality, NYHA functional class and echocardiographic measurements were followed up at 30 days and at 3, 9 and 12 months postoperatively.

RESULTS

Overall, mean age was 79 ± 9 years, mean left ventricular ejection fraction 50 ± 17% and mean log EuroSCORE 28.3 ± 17.1%. There were no perioperative complications. Paravalvular regurgitation immediately after implantation was graded none (n = 6), trace (n = 3) or mild (n = 1). Overall 30-day mortality was 30% (3/10). Three patients refused further treatment, such as haemodialysis or treatment of mitral regurgitation. Rate for pacemaker implantation was 2/10 (20%).

CONCLUSIONS

Intraprocedural success and haemodynamic data in our cases were good. The mortality in our group highlighted the importance of careful patient selection, especially for this pathology. The Jenavalve prosthesis proved to be suitable for treatment of AR in surgical high-risk patients.

Для цитирования: Schlingloff Friederike, Schäfer Ulrich, Frerker Christian, Schmoeckel Michael, Bader Ralf. Transcatheter aortic valve implantation of a second-generation valve for pure aortic regurgitation: procedural outcome, haemodynamic data and follow-up.. Interactive cardiovascular and thoracic surgery. 2014. DOI: 10.1093/icvts/ivu155.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.