m+Medelevator.RU
← Поиск публикаций
Annals of internal medicine · 2023

Implantable Defibrillator System Shock Function, Mortality, and Cause of Death After Magnetic Resonance Imaging.

Клиническое исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ra Joshua, Oberdier Matt T, Suzuki Masahito, Vaidya Dhananjay, Liu Yisi, Hansford Rozann, McVeigh Diana, Weltin Valeria, Tao Susumu, Thiemann David R, Nazarian Saman, Halperin Henry R

Аннотация

BACKGROUND: Studies have shown that magnetic resonance imaging (MRI) does not have clinically important effects on the device parameters of non-MRI-conditional implantable cardioverter-defibrillators (ICDs). However, data on non-MRI-conditional ICD detection and treatment of arrhythmias after MRI are limited.

OBJECTIVE: To examine if non-MRI-conditional ICDs have preserved shock function of arrhythmias after MRI.

DESIGN: Prospective cohort study. (ClinicalTrials.gov: NCT01130896).

SETTING: 1 center in the United States.

PATIENTS: 629 patients with non-MRI-conditional ICDs enrolled consecutively between February 2003 and January 2015.

INTERVENTIONS: 813 total MRI examinations at a magnetic field strength of 1.5 Tesla using a prespecified safety protocol.

MEASUREMENTS: Implantable cardioverter-defibrillator interrogations were collected after MRI. Clinical outcomes included arrhythmia detection and treatment, generator or lead exchanges, adverse events, and death.

RESULTS: During a median follow-up of 2.2 years from MRI to latest available ICD interrogation before generator or lead exchange in 536 patients, 4177 arrhythmia episodes were detected, and 97 patients received ICD shocks. Sixty-one patients (10% of total) had 130 spontaneous ventricular tachycardia or fibrillation events terminated by ICD shocks. A total of 210 patients (33% of total) are known to have died (median, 1.7 years from MRI to death); 3 had cardiac arrhythmia deaths where shocks were indicated without direct evidence of device dysfunction.

LIMITATIONS: Data were acquired at a single center and may not be generalizable to other clinical settings and MRI facilities. Implantable cardioverter-defibrillator interrogations were not available for a subset of patients; adjudication of cause of death relied solely on death certificate data in a subset.

CONCLUSION: Non-MRI-conditional ICDs appropriately treated detected tachyarrhythmias after MRI. No serious adverse effects on device function were reported after MRI.

PRIMARY FUNDING SOURCE: Johns Hopkins University and National Institutes of Health.