Usability and Accuracy of a Novel Automatic Bladder Diary Device (Uro-diary Checker).
СтатьяМногоцентровое исследованиеНаблюдательное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Okumura Yoshinaga, Amaya Yoshie, Inamura Sou, Inamura Mahiro, Kaeriyama Kimika, Okubo Nodoka, Shimada Sahoko, Kakitsuba Tadashi, Seo Wonseok, Tanaka Nobuki, Fukiage Yusuke, Kabuto Takafumi, Tsutsumiuchi Manami, Seki Masaya, Taga Minekatsu, Terada Naoki
Аннотация
OBJECTIVE: To evaluate the usability and accuracy of a newly developed automated bladder diary (BD) device, the Uro-diary Checker (UDC), in comparison with a conventional handwritten BD.
METHODS: In this prospective, multi-institutional observational study, 40 patients with lower urinary tract symptoms (LUTS) were recruited between April 2023 and October 2024. Participants recorded their voiding patterns using a handwritten BD for 3 days and the UDC for the next 3 days. The UDC automatically recorded voiding time and volume when a urine cup was placed on the device, and the data were transferred via Bluetooth to a computer. The primary endpoint was usability, assessed using the System Usability Scale (SUS), a validated 10-item Likert-scale questionnaire.
RESULTS: Of the 40 enrolled patients, 31 successfully completed records using both methods and were included in the final analysis. The mean patient age was 72.6 ± 9.8 years; 24 were male and 7 were female. The SUS score was significantly higher for the UDC (72.9 ± 19.3) than for the BD (60.6 ± 19.5; P < .01), exceeding the benchmark score of 68 and indicating good usability. UDC-recorded urine volumes were highly consistent with BD measurements for both 24-hour urine volume (r = 0.86, P < .01) and average voided volume (r = 0.77, P < .01).
CONCLUSION: The UDC demonstrated significantly greater usability than the conventional BD while maintaining comparable accuracy for urine volume measurements, suggesting its potential to improve adherence and efficiency in LUTS evaluation.
METHODS: In this prospective, multi-institutional observational study, 40 patients with lower urinary tract symptoms (LUTS) were recruited between April 2023 and October 2024. Participants recorded their voiding patterns using a handwritten BD for 3 days and the UDC for the next 3 days. The UDC automatically recorded voiding time and volume when a urine cup was placed on the device, and the data were transferred via Bluetooth to a computer. The primary endpoint was usability, assessed using the System Usability Scale (SUS), a validated 10-item Likert-scale questionnaire.
RESULTS: Of the 40 enrolled patients, 31 successfully completed records using both methods and were included in the final analysis. The mean patient age was 72.6 ± 9.8 years; 24 were male and 7 were female. The SUS score was significantly higher for the UDC (72.9 ± 19.3) than for the BD (60.6 ± 19.5; P < .01), exceeding the benchmark score of 68 and indicating good usability. UDC-recorded urine volumes were highly consistent with BD measurements for both 24-hour urine volume (r = 0.86, P < .01) and average voided volume (r = 0.77, P < .01).
CONCLUSION: The UDC demonstrated significantly greater usability than the conventional BD while maintaining comparable accuracy for urine volume measurements, suggesting its potential to improve adherence and efficiency in LUTS evaluation.