Journal of vascular and interventional radiology : JVIR · 2004

CT and MR imaging of nitinol stents with radiopaque distal markers.

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Létourneau-Guillon Laurent, Soulez Gilles, Beaudoin Gilles, Oliva Vincent L, Giroux Marie-France, Qin Zhao, Boussion Nicolas, Therasse Eric, De Guise Jacques, Cloutier Guy

Аннотация

PURPOSE: To evaluate imaging characteristics and artifacts of a nitinol stent with distal tantalum markers with computed tomography (CT) angiography and magnetic resonance (MR) angiography.

MATERIALS AND METHODS: A vascular phantom was built to simulate in-stent restenosis. A nitinol stent with tantalum markers (Luminexx stent) was evaluated with CT angiography in different orientations relative to the z-axis and with MR angiography in different positions relative to both B0 and the readout gradient. Stenosis measurements were compared with conventional digital subtraction angiography for both modalities. In-stent signal intensity obtained with different flip angles was assessed in two nitinol stents with distal markers (Luminexx stent and SMART stent) and one without markers (Memotherm-FLEXX stent).

RESULTS: Stenosis detection was not possible with CT angiography when the stent was perpendicular to the z-axis because of streak-like artifacts induced by tantalum markers. Stenosis evaluation with multiplanar reformation was accurate when the stent was in parallel and oblique orientations relative to the table axis. With MR angiography, metallic artifacts were mostly related to the stent orientation with B0, whereas orientation of the readout gradient had little influence. The mean error (overestimation) for stenosis measurements varied between 0.1% and 7.4% for CT imaging in parallel and oblique positions and 3.6% and 9.5% for MR imaging. Higher flip angles did not improve signal intensity inside the three stents tested.

CONCLUSION: CT and MR angiography can be used for evaluating the patency of stents with distal markers that are parallel or oblique relative to the table axis (iliac, carotid, or femoral stents). MR angiography is preferred if the stent is perpendicular to the table axis (renal stent).