Flash glucose monitoring system in patients with type 1 diabetes in healthcare center in Brazil: real world data from a short-term prospective study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Matheus Alessandra S M, Pascoal Juliana Brazil Fontes, Cabizuca Carolina Alves, Tannus Lucianne Righeti Monteiro, Guimarães Rafael Seno, Mattos Diogo Menezes Ferrazani, Cobas Roberta Arnoldi
Аннотация
OBJECTIVE: To evaluate the alternate use of flash glucose monitoring (FGM) with self-monitoring blood glucose (SMBG), in patients with type 1 diabetes (T1D).
MATERIALS AND METHODS: Two weeks of open FGM (P2), both preceded (P1) and followed by 2 weeks (P3) of SMBG with a blinded FGM system. Mean absolute relative difference (MARD) was calculated by (|FGMi - SMBGi|) / SMBGi, where it was a paired data sample.
RESULTS: In total, 34 patients were evaluated. Time in range (TIR) did not change between P1 and P2. In 12 patients (35.3%), TIR increased from 40% at P1 to 52% at P2 (p = 0.002) and in 22 (64.7%), TIR decreased or did not change. FGM use resulted in decreased % time spent in hypoglycemia (<70 mg/dL) (6.5% vs. 5.0%; p = 0.005), increased % time spent in hyperglycemia (>180 mg/dL) (44.5% to 51%; p = 0.046) with no significant change in % TIR. The proportion of patients who reached sensor-estimated glycated hemoglobin (eA1c) < 7% decreased from 23.5% at P1 to 12.9% at P2, p = 0.028. For the whole sample, the MARD between the two methods was 15.5% (95% CI 14.5-16.6%). For normal glucose range, hyperglycemic levels and hypoglycemic levels MARD were 16.0% (95% CI 15.0-17.0%), 13.3% (95% CI 11.5-15.2%) and 23.4% [95% CI 20.5-26.3%)], respectively.
CONCLUSION: FGM after usual SMBG decreased the % time spent in hypoglycemia concomitant with an undesired increase in % time spent in hyperglycemia. Lower accuracy of FGM regarding hypoglycemia levels could result in overcorrection of hypoglycemia.
MATERIALS AND METHODS: Two weeks of open FGM (P2), both preceded (P1) and followed by 2 weeks (P3) of SMBG with a blinded FGM system. Mean absolute relative difference (MARD) was calculated by (|FGMi - SMBGi|) / SMBGi, where it was a paired data sample.
RESULTS: In total, 34 patients were evaluated. Time in range (TIR) did not change between P1 and P2. In 12 patients (35.3%), TIR increased from 40% at P1 to 52% at P2 (p = 0.002) and in 22 (64.7%), TIR decreased or did not change. FGM use resulted in decreased % time spent in hypoglycemia (<70 mg/dL) (6.5% vs. 5.0%; p = 0.005), increased % time spent in hyperglycemia (>180 mg/dL) (44.5% to 51%; p = 0.046) with no significant change in % TIR. The proportion of patients who reached sensor-estimated glycated hemoglobin (eA1c) < 7% decreased from 23.5% at P1 to 12.9% at P2, p = 0.028. For the whole sample, the MARD between the two methods was 15.5% (95% CI 14.5-16.6%). For normal glucose range, hyperglycemic levels and hypoglycemic levels MARD were 16.0% (95% CI 15.0-17.0%), 13.3% (95% CI 11.5-15.2%) and 23.4% [95% CI 20.5-26.3%)], respectively.
CONCLUSION: FGM after usual SMBG decreased the % time spent in hypoglycemia concomitant with an undesired increase in % time spent in hyperglycemia. Lower accuracy of FGM regarding hypoglycemia levels could result in overcorrection of hypoglycemia.