Accuracy of Flash Glucose Monitoring in Hemodialysis Patients With and Without Diabetes Mellitus.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Weber Michèle R, Diebold Matthias, Wiesli Peter, Kistler Andreas D

Аннотация

AIMS: Glucose and insulin metabolism are altered in hemodialysis patients, and diabetes management is difficult in these patients. We aimed to validate flash glucose monitoring (FGM) in hemodialysis patients with and without diabetes mellitus as an attractive option for glucose monitoring not requiring regular self-punctures.

METHODS: We measured interstitial glucose using a FreeStyle Libre device in eight hemodialysis patients with and seven without diabetes mellitus over 14 days and compared the results to simultaneously performed self-monitoring of capillary blood glucose (SMBG).

RESULTS: In 720 paired measurements, mean flash glucose values were significantly lower than self-measured capillary values (6.17±2.52 vs. 7.15±2.41 mmol/L, p=1.3 E-86). Overall, the mean absolute relative difference was 17.4%, and the mean absolute difference was 1.20 mmol/L. The systematic error was significantly larger in patients without vs. with diabetes (- 1.17 vs. - 0.82 mmol/L) and on dialysis vs. interdialytic days (-1.09 vs. -0.90 mmol/L). Compared to venous blood glucose (72 paired measurements), the systematic error of FGM was even larger (5.89±2.44 mmol/L vs. 7.78±7.25 mmol/L, p=3.74E-22). Several strategies to reduce the systematic error were evaluated, including the addition of +1.0 mmol/L as a correction term to all FGM values, which significantly improved accuracy.

CONCLUSIONS: FGM systematically underestimates blood glucose in hemodialysis patients but, taking this systematic error into account, the system may be useful for glucose monitoring in hemodialysis patients with or without diabetes.