Clinical Evaluation of Assays for Plasma Renin Activity and Aldosterone Measurement by Liquid Chromatography-Tandem Mass Spectrometry.

Gibbons Stephen M, Field Helen P, Fairhurst April, Fleming Anne, Ford Clive, Williams Emma L, Barnes Sophie C, Barth Julian H

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Aldosterone and renin are pivotal hormones in the regulation of salt and water homeostasis and blood pressure. Measurement of renin and aldosterone in serum/plasma is essential for the investigation of primary hyperaldosteronism (PA) and monitoring of glucocorticoid replacement therapy.

METHODS

We report 2 LC-MS/MS methods developed to measure aldosterone and plasma renin activity (PRA). PRA was determined by endogenous enzymatic generation of angiotensin I using 150 µL of sample. Generated angiotensin I was purified by solid phase extraction prior to chromatographic separation and mass spectrometry. Aldosterone measurement required 300 μL of sample extracted with MTBE prior to LC-MS/MS analysis.

RESULTS

The PRA method was linear (1.2-193 nmol/L), sensitive (LLOQ = 1.2 nmol/L), precise (CV = 4.1%), and specific (no cross reactivity for a number of structurally similar steroids). Dilutional linearity and recovery (84%) were acceptable. Accuracy was confirmed by comparison against our current RIA method. The aldosterone method had equally acceptable performance characteristics. Reference ranges in 110 healthy normotensive subjects were: PRA 0.2-3.7 nmol/L/h and aldosterone 50-950 pmol/L. Consecutive patients (n = 62) with adrenal incidentalomas shown to have no functional adrenal disease; their post overnight 1 mg dexamethasone test values were: PRA 0.2-2.6 nmol/L/h and aldosterone 55-480 pmol/L. Serum aldosterone values after 2 liter saline suppression were-normal subjects (n = 17): 78-238 pmol/L and confirmed primary hyperaldosteronism (n = 25): 131-1080 pmol/L.

CONCLUSIONS

We have developed robust assays for PRA and aldosterone with appropriate clinical evaluation. These assays are now in routine practice in the UK.

Для цитирования: Gibbons Stephen M, Field Helen P, Fairhurst April, Fleming Anne, Ford Clive, Williams Emma L, Barnes Sophie C, Barth Julian H. Clinical Evaluation of Assays for Plasma Renin Activity and Aldosterone Measurement by Liquid Chromatography-Tandem Mass Spectrometry.. The journal of applied laboratory medicine. 2021. DOI: 10.1093/jalm/jfaa177.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.