MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/neurintsurg-2013-010710.rep

Медицинская библиотека · Journal of neurointerventional surgery · 2013

Entrapment of a micro-guidewire during stenting of basilar stenosis.

Ma Ning, Liu Lian, Wang Tie-Jun, Xu Xiao-Tong, Miao Zhong-Rong

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Entrapment of a microwire related to intracranial endovascular therapy is an uncommon complication. A 64-year-old man with symptomatic basilar artery stenosis was treated with stenting. A 300 cm Synchro microwire became trapped in the P1 segment of the left posterior cerebral artery during the procedure. The decision was made to leave the microwire in place, considering that aggressive retrieval procedures could cause injury to the cerebral vasculature. The entrapped microwire was later inadvertently dislocated and removed while pulling the microwire back into the femoral artery with a looped catheter from the left femoral access. This demonstrates that, although entrapment of a microwire during endovascular therapy is a rare event, an optimal salvage technique needs to be explored further.

Для цитирования: Ma Ning, Liu Lian, Wang Tie-Jun, Xu Xiao-Tong, Miao Zhong-Rong. Entrapment of a micro-guidewire during stenting of basilar stenosis.. Journal of neurointerventional surgery. 2013. DOI: 10.1136/neurintsurg-2013-010710.rep.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.