DOI 10.1111/j.1540-8159.2005.00157.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2005

A technique of lead insertion for low atrial septal pacing.

De Voogt Willem G, Van Mechelen Rob, Van Den Bos Arjan, Scheffer Mike, Van Hemel Norbert M, Koistinen Juhani

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Pacing in the low right atrial septum (LAS) appears superior to right atrial appendage or free wall stimulation for the prevention of paroxysmal atrial fibrillation. However, insertion of active fixation lead in the low right atrial septal position is difficult and time consuming, inhibiting application of this pacing method in daily practice.

METHODS

The technique of handling and positioning of a new "over the wire" lead system is presented with emphasis on electrocardiographic P wave pattern and fluoroscopic landmarks.

RESULTS

The initial results demonstrate an acute implantation and short-term success of LAS pacing of >90% in the first 100 patients without major complications. Pacing thresholds at 3 and 6 months were fully comparable with that of the conventional atrial pacing, whereas impedance and atrial sensing signals were significantly higher at 3 and 6 months follow-up.

CONCLUSION

These favorable initial results justify recommanding chronic LAS pacing with the active fixation atrial lead and providinge guidelines and fluoroscopic landmarks for the implantation. Insertion of the atrial active fixation lead positioned with the Locatortrade mark tool strongly supports the implantation procedure.

Для цитирования: De Voogt Willem G, Van Mechelen Rob, Van Den Bos Arjan, Scheffer Mike, Van Hemel Norbert M, Koistinen Juhani. A technique of lead insertion for low atrial septal pacing.. Pacing and clinical electrophysiology : PACE. 2005. DOI: 10.1111/j.1540-8159.2005.00157.x.