Ultramicro-fluorometric assay for the diagnosis of Gaucher disease in dried blood spots on filter paper.
Аннотация
BACKGROUND
Gaucher disease (GD) is a lysosomal storage disorder characterized by a deficiency of the lysosomal acid β-D-glucosidase (GBA). The aim of this study was to develop an ultramicro-fluorometric assay based on the method of Chamoles et al. for determining GBA activity in dried blood spots on filter paper (DBS).
METHODS
The assay used 3-mm diameter blood spot and 8 mmol/l of 4-methylumbelliferyl-β-D-glucoside as enzymatic substrate. The reaction occurred in plates incubated at 37°C for 20 hours and the enzyme activity was expressed in μmol hydrolysed substrate/l blood/h. The fluorescence of the enzyme product was automatically measured in a fluorometer-photometer reader (SUMA Technology).
RESULTS
The intra and inter-assay coefficients of variation were lower than 9 and 12%, respectively, and the recovery range was 97-109%.Three patients with GD were correctly diagnosed using the ultramicroassay. Healthy newborn DBS samples (n = 3003) from the National Neonatal Screening Program were analyzed, and the mean GBA activity was 5.7 μmol/l blood/h. Our assay showed high Pearson (n = 26; r = 0.99) and concordance correlations (ρc = 0.99) with the traditional method described by Chamoles et al.
CONCLUSIONS
The analytical performance characteristics of our ultramicro-fluorometric assay suggest that it can be used in the diagnosis of GD in newborns and adults.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.