A novel design biodegradable stent for use in congenital heart disease: mid-term results in rabbit descending aorta.

Veeram Reddy Surendranath R, Welch Tré R, Wang Jian, Richardson James A, Forbess Joseph M, Riegel Matthew, Nugent Alan W

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study evaluates the feasibility of delivery and deployment of low and medium molecular weight (LMW and MMW, respectively) double-opposing helical (DH) poly-l-lactic acid biodegradable stent (BDS) in rabbit descending aorta (DAO). Secondary objectives were to assess patency and inflammation of stented vessels at 9 months and to investigate safety following intentional embolization of stent fragments in DAO.

BACKGROUND

A BDS that will relieve aortic obstruction and disappears as the child grows older allowing for preservation of aortic wall elasticity and natural growth of aorta will be ideal to treat Coarctation (CoA). BDS have never been evaluated in the DAO.

METHODS

Seven New Zealand white rabbits underwent implantation of DH-LMW (n = 7), DH-MMW (n = 3), and metal stents (n = 7) in DAO. BDS fragments were intentionally embolized into DAO in two rabbits.

RESULTS

All stents were deployed via a 6-French sheath. Five BDS covered the origin of major DAO side branches. Angiography and intravascular ultrasound showed good stent apposition to the wall of DAO with minimal luminal loss at 9 months follow-up. All stents had minimal neointimal hyperplasia on histopathology. Adverse events included 1 death, 1 aortic aneurysm, and lower extremity ulceration due to self-mutilation in an embolization rabbit.

CONCLUSIONS

Pilot study confirms the feasibility of delivery and deployment of up to 6-millimeter diameter DH BDS in rabbit DAO. Stent integrity with DH design was maintained at 9 months with minimal vessel inflammation. Potential morbidity due to embolized BD fragments cannot be ruled out and needs further evaluation.

Для цитирования: Veeram Reddy Surendranath R, Welch Tré R, Wang Jian, Richardson James A, Forbess Joseph M, Riegel Matthew, Nugent Alan W. A novel design biodegradable stent for use in congenital heart disease: mid-term results in rabbit descending aorta.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2014. DOI: 10.1002/ccd.25648.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.