DOI 10.1016/j.jcin.2024.08.050

Медицинская библиотека · JACC. Cardiovascular interventions · 2025

Early and Late Outcomes With the Absorb Bioresorbable Vascular Scaffold: Final Report From the ABSORB Clinical Trial Program.

Power David A, Camaj Anton, Kereiakes Dean J, Ellis Stephen G, Gao Runlin, Kimura Takeshi, Ali Ziad A, Stockelman Kelly A, Dressler Ovidiu, Onuma Yoshinobu, Serruys Patrick W, Stone Gregg W, ABSORB Investigators

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The risk-benefit ratio of the Absorb bioresorbable vascular scaffold (BVS) may vary before and after 3 years, the time point of complete bioresorption of the poly-L-lactic acid scaffold.

OBJECTIVES

The aim of this study was to determine the time-varying outcomes of the Absorb BVS compared with cobalt-chromium everolimus-eluting stents (EES) from a large individual-patient-data pooled analysis of randomized trials.

METHODS

The individual patient data from 5 trials that randomized 5,988 patients undergoing percutaneous coronary intervention to the Absorb BVS vs EES with 5-year follow-up were pooled. The primary effectiveness and safety endpoints were target lesion failure (TLF) (cardiac death, target vessel myocardial infarction, or ischemia-driven target lesion revascularization) and device thrombosis (DT).

RESULTS

Between 0 and 5 years, TLF occurred in 15.9% BVS patients vs 13.1% EES patients (HR: 1.25; 95% CI: 1.08-1.43; P = 0.002), and DT occurred in 2.2% vs 1.0%, respectively (HR: 2.38; 95% CI: 1.49-3.79; P = 0.0002). Between 0 and 3 years, TLF occurred in 12.4% BVS patients vs 9.3% EES patients (HR: 1.35; 95% CI: 1.15-1.59; P = 0.0002), and DT occurred in 2.0% vs 0.6%, respectively (HR: 3.58; 95% CI: 2.01-6.36; P < 0.0001). Between 3 and 5 years, TLF occurred in 4.5% BVS patients vs 4.7% EES patients (HR: 0.99; 95% CI: 0.76-1.27; P = 0.91), and DT occurred in 0.2% vs 0.4%, respectively (HR: 0.49; 95% CI: 0.18-1.38; P = 0.17). By spline analysis, the TLF hazard by 5 years was lower with BVS than EES.

CONCLUSIONS

Compared with EES treatment, BVS increased TLF and DT rates through 5-year follow-up. However, the period of excess risk for the first-generation Absorb BVS ended at 3 years, when poly-L-lactic acid bioresorption is complete. Thereafter event rates were comparable or lower with BVS. (ABSORB II Randomized Controlled Trial, NCT01425281; A Clinical Evaluation of Absorb™ Bioresorbable Vascular Scaffold [Absorb™ BVS] System in Chinese Population - ABSORB CHINA Randomized Controlled Trial [ABSORB CHINA], NCT01923740; AVJ-301 Clinical Trial: A Clinical Evaluation of AVJ-301 [Absorb™ BVS] in Japanese Population [ABSORB JAPAN], NCT01844284; ABSORB III Randomized Controlled Trial [ABSORB III], NCT01751906; Absorb IV Randomized Controlled Trial, NCT02173379).

Для цитирования: Power David A, Camaj Anton, Kereiakes Dean J, Ellis Stephen G, Gao Runlin, Kimura Takeshi, Ali Ziad A, Stockelman Kelly A, Dressler Ovidiu, Onuma Yoshinobu, Serruys Patrick W, Stone Gregg W, ABSORB Investigators. Early and Late Outcomes With the Absorb Bioresorbable Vascular Scaffold: Final Report From the ABSORB Clinical Trial Program.. JACC. Cardiovascular interventions. 2025. DOI: 10.1016/j.jcin.2024.08.050.