DOI 10.1007/s00117-026-01653-z
Feasibility of reducing radiation dose and contrast dose while improving image quality in dual-energy CT pulmonary angiography : Use of low-energy virtual monochromatic images with deep learning image reconstruction algorithm.
Аннотация
BACKGROUND
We explored the feasibility of reducing both radiation and contrast doses while improving image quality using low-energy virtual monochromatic images (VMIs) in dual-energy computed tomography (CT) pulmonary angiography (DECTPA) with deep learning image reconstruction at a high setting (DLIR-H).
MATERIALS AND
METHODS
A total of 60 patients scheduled for CTPA were randomly divided into two groups: group A (n = 30) with 120 kV, contrast dose of 0.8 mL/kg, and adaptive statistical iterative reconstruction‑V (ASIR-V) at a 60% strength level, and group B (n = 30) with dual-energy CT (DECT), contrast dose of 0.6 mL/kg, and DLIR‑H at 40 keV, 50 keV, 60 keV and 70 keV VMIs. The CT and standard deviation (SD) values of the pulmonary arteries were measured to calculate the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Image quality was subjectively scored blindly by two radiologists using a five-point scale.
RESULTS
There were no differences in general patient demographics between the two groups (all p > 0.05). In group B, the contrast and radiation doses decreased by 29.6% and 53.1%, respectively, compared to group A (p < 0.05). Compared to group A, group B had significantly higher CT and SNR values for the 40 keV-60 keV VMIs and higher CNR for the 40 keV-70 keV VMIs (all p < 0.05), with similar image noise at 60 keV. The two radiologists had substantial agreement in subjective scoring (kappa > 0.80). There were significantly higher scores for the 40 keV-60 keV VMIs in group B compared to group A (all p < 0.05), with the 40 keV VMIs having the highest scores.
CONCLUSION
Combining DLIR‑H with VMIs, especially 40 keV, in DECTPA leads to reduced contrast and radiation doses while improving image quality.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.