DOI 10.1097/00126097-200402000-00003

Медицинская библиотека · Blood pressure monitoring · 2004

Venous blood redistribution alters the accuracy of Omron HEM-705CP.

Czarkowski Marek, Mikulska Małgorzata, ZEbrowski Mariusz, Baran Anna, Rózanowski Krzysztof

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Validation procedures of automated blood pressure measuring devices are carried out at rest only.

DESIGN

We aimed to determine whether venous blood redistribution affects the accuracy of the OMRON HEM-705CP (Osc). Lower body negative pressure stimulus (-40 mmHg) (LBNP) was used to simulate changes of body position.

METHODS

Fifty-four young healthy volunteers had their blood pressure (BP) measured simultaneously by Osc and mercury sphygmomanometer (HgS) at rest (at 3 and 5 min) and during LBNP (at 7 and 9 min).

RESULTS

Differences (delta) between HgS and Osc for systolic and diastolic BP (SBP and DBP) were calculated. The deltaSBP obtained in minute 9 (deltaSBP-9) was bigger than the one taken in minute 5(deltaSBP-5) (-2.89+/- 4.3 versus -1.35+/-5.2 mmHg, P<0.02). The rest of the deltaSBP values and all the deltaDBP values were constant at rest and during venous blood redistribution caused by LBNP. We calculated the fall in SBP separately for HgS and Osc by subtracting SBP-9 from SBP-5 for each patient. The fall measured by Osc was bigger than by HgS (-5.87+/-4.0 versus -4.3+/-4.5 mmHg; P<0.02). When graded according to British Hypertension Society protocol, Osc remained in the class of recommendation (A or B) for every minute of the procedure.

CONCLUSIONS

Even if an oscillometric device passes various validation tests, they do not give sufficient information about its accuracy during venous blood redistribution. This knowledge is of particular importance when BP regulation abnormalities occurring during changes in body position are diagnosed with the oscillometric method.

Для цитирования: Czarkowski Marek, Mikulska Małgorzata, ZEbrowski Mariusz, Baran Anna, Rózanowski Krzysztof. Venous blood redistribution alters the accuracy of Omron HEM-705CP.. Blood pressure monitoring. 2004. DOI: 10.1097/00126097-200402000-00003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.