Bioengineering (Basel, Switzerland) · 2023
Exerting Forces and Wall Load during Duodenoscopy for ERCP: An Experimental Measurement in an Artificial Model.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Schneider Julian, Duckworth-Mothes Benedikt, Schweizer Ulrich, Königsrainer Alfred, Fisch Jakob, Wichmann Dörte
Аннотация
BACKGROUND: Endoscopic retrograde cholangiopancreatography (ERCP) is crucial to the treatment of biliopancreatic diseases with iatrogenic perforation as a potential complication. As of yet, the wall load during ERCP is unknown, as it is not directly measurable during an ERCP in patients.
METHODS: In a life-like, animal-free model, a sensor system consisting of five load cells was attached to the artificial intestines (sensors 1 + 2: pyloric canal-pyloric antrum, sensor 3: duodenal bulb, sensor 4: descending part of the duodenum, sensor 5: distal to the papilla). Measurements were made with five duodenoscopes (n = 4 reusable and n = 1 single use).
RESULTS: Fifteen standardized duodenoscopies were performed. Peak stresses were found at the antrum during the gastrointestinal transit (sensor 1 max. 8.95 N, sensor 2 max. 2.79 N). The load reduced from the proximal to the distal duodenum and the greatest load in the duodenum was discovered at the level of the papilla in 80.0% (sensor 3 max. 2.06 N).
CONCLUSIONS: For the first time, intraprocedural load measurements and exerting forces obtained during a duodenoscopy for ERCP in an artificial model were recorded. None of the tested duodenoscopes were classified as dangerous for patient safety.
METHODS: In a life-like, animal-free model, a sensor system consisting of five load cells was attached to the artificial intestines (sensors 1 + 2: pyloric canal-pyloric antrum, sensor 3: duodenal bulb, sensor 4: descending part of the duodenum, sensor 5: distal to the papilla). Measurements were made with five duodenoscopes (n = 4 reusable and n = 1 single use).
RESULTS: Fifteen standardized duodenoscopies were performed. Peak stresses were found at the antrum during the gastrointestinal transit (sensor 1 max. 8.95 N, sensor 2 max. 2.79 N). The load reduced from the proximal to the distal duodenum and the greatest load in the duodenum was discovered at the level of the papilla in 80.0% (sensor 3 max. 2.06 N).
CONCLUSIONS: For the first time, intraprocedural load measurements and exerting forces obtained during a duodenoscopy for ERCP in an artificial model were recorded. None of the tested duodenoscopes were classified as dangerous for patient safety.