m+Medelevator.RU
← Поиск публикаций
The Journal of arthroplasty · 2019

High Survival Rate and Very Low Wear of Lateral Unicompartmental Arthroplasty at Long Term: A Case Series of 54 Cases at a Mean Follow-Up of 17 Years.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Deroche Etienne, Batailler Cécile, Lording Timothy, Neyret Philippe, Servien Elvire, Lustig Sébastien

Аннотация

BACKGROUND: Survivorship of lateral unicompartmental knee arthroplasty (UKA) has progressively improved. However, there are few studies describing long-term results, and no study reports on polyethylene (PE) wear in lateral unicompartmental arthroplasty. The aims of this study are to determine the survival rate of lateral UKA with a fixed, all-PE bearing, and the PE wear of the tibial implant at a minimum of 15 years follow-up.

METHODS: From January 1988 to October 2003, we performed 54 lateral UKAs in 52 patients. All patients had isolated lateral osteoarthritis (OA). The mean age at the index procedure was 65.4 ± 11 years. Thirty-nine UKAs were available for follow-up (30 alive and 9 dead after 15 years). Twelve patients had died before 15 years and 3 patients were lost to follow-up. The mean follow-up was 17.9 years (range, 15-23 years).

RESULTS: At the final follow-up, 8 knees of 39 (20.5%) had a surgical revision. The cumulative survival rate was 82.1% at 15 years and 79.4% at 20 years. The main reason of revision was progression of OA (87.5%), followed by aseptic loosening of the tibial component (12.5%). With a mean follow-up of 17.9 years, the mean PE wear was 0.061 mm/y. There was no radiographic loosening in the surviving implants and no revisions for wear. The mean functional International Knee Society score was 66.5 ± 26.8, with a mean objective score of 84.4 points ± 13.2. In the population without revision, 90.5% were satisfied or very satisfied at the latest follow-up.

CONCLUSION: Lateral UKA with a fixed, all-PE tibial bearing and a femoral resurfacing implant presents a high survivorship at long term, with very low PE wear.