Does the disinfection of a photostimulable phosphor plate receptor with peracetic acid impair carious lesions diagnosis?
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ruiz Débora Costa, Haiter-Neto Francisco, Gaêta-Araujo Hugo, Freitas Deborah Queiroz
Аннотация
OBJECTIVES: To investigate the impact of disinfecting a photostimulable phosphor plate (PSP) receptor with peracetic acid on proximal carious lesions diagnosis. Additionally, to evaluate and measure artefacts caused by disinfection.
METHODS: The proximal surfaces of 40 posterior teeth were evaluated by microcomputed tomography to confirm whether they were sound or carious. These teeth were then mounted in 20 phantoms. Using a PSP receptor (Express System), 1 radiograph of each phantom was acquired. Then, 20 disinfections with peracetic acid were performed, corresponding to a disinfection cycle. After the first disinfection cycle, new radiographs were acquired. This process was repeated 9 times, totalling 200 disinfections (10 cycles × 20 disinfections) and 220 radiographs (200 radiographs + 20 baseline radiographs). The radiographs were assessed independently by 5 examiners for carious lesions diagnosis. A radiologist identified and measured the size of the artefacts. Area under the receiver-operating characteristic curve, sensitivity, and specificity were calculated and compared among the number of disinfections by analysis of variance (α = 0.05). The weighted Kappa index evaluated intra- and inter-examiner agreements.
RESULTS: The disinfections did not affect the diagnostic metrics for detecting proximal carious lesions (P > .05). After 200 disinfections, the biggest artefact measured 0.24 cm. Intra- and inter-examiner agreements ranged from moderate to substantial (0.60-0.69) and from fair to moderate (0.34-0.60).
CONCLUSIONS: After 200 disinfections, the radiographic detection of proximal carious lesions remains unaffected by the disinfection with peracetic acid, and the artefacts are still small.
METHODS: The proximal surfaces of 40 posterior teeth were evaluated by microcomputed tomography to confirm whether they were sound or carious. These teeth were then mounted in 20 phantoms. Using a PSP receptor (Express System), 1 radiograph of each phantom was acquired. Then, 20 disinfections with peracetic acid were performed, corresponding to a disinfection cycle. After the first disinfection cycle, new radiographs were acquired. This process was repeated 9 times, totalling 200 disinfections (10 cycles × 20 disinfections) and 220 radiographs (200 radiographs + 20 baseline radiographs). The radiographs were assessed independently by 5 examiners for carious lesions diagnosis. A radiologist identified and measured the size of the artefacts. Area under the receiver-operating characteristic curve, sensitivity, and specificity were calculated and compared among the number of disinfections by analysis of variance (α = 0.05). The weighted Kappa index evaluated intra- and inter-examiner agreements.
RESULTS: The disinfections did not affect the diagnostic metrics for detecting proximal carious lesions (P > .05). After 200 disinfections, the biggest artefact measured 0.24 cm. Intra- and inter-examiner agreements ranged from moderate to substantial (0.60-0.69) and from fair to moderate (0.34-0.60).
CONCLUSIONS: After 200 disinfections, the radiographic detection of proximal carious lesions remains unaffected by the disinfection with peracetic acid, and the artefacts are still small.