DOI 10.1016/j.ejrad.2016.03.005

Медицинская библиотека · European journal of radiology · 2016

Assessment of right atrium volume by conventional CT or MR techniques: Which modality resembles in vivo reality?

Rheinheimer Stephan, Reh Christina, Figiel Jens, Mahnken Andreas H

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Right atrial volume (RAV) is a prognostic factor of the right heart function. This retrospective study evaluates the comparability of computed tomography (CT) and magnetic resonance imaging (MRI) for right atrial volumetry with conventional planimetric and diametric methods.

MATERIAL AND

METHODS

29 retrospectively included patients (18 male, 47±12years) underwent CT and 1.5 Tesla MRI within 17±20days. RAV was measured using biplane and ellipsoid method (MRI and CT) and Simpson's method (CT). For interobserver comparison measurements were carried out by two observers. Pearson's correlation, Lin's concordance coefficient, and Bland-Altman statistics were calculated.

RESULTS

There is a correlation of RAV between CT and MRI, r [ellipsoid]=0.65; r [biplane]=0.64 (p<0.001). MRI volumes were significantly lower than CT volumes, [mean±SD] 43±19ml versus 27±9ml, (p<0.002). There was a close interobserver correlation (CT: r=0.83, MRI: r=0.73; p<0.001) but a relatively wide range in Bland-Altman analysis; limits of agreement from ±13ml up to ±29ml.

CONCLUSIONS

Compared to left atrial volumes, a higher variability was found for RAV values both in interobserver statistiscs and in intermodality comparison. Complex shape of the right atrium, artifacts due to contrast material (CT), high venous return in inspiration (CT), high flow contrast media administration (CT) and increased heart rate (MRI) might cause these clinically relevant variations.

Для цитирования: Rheinheimer Stephan, Reh Christina, Figiel Jens, Mahnken Andreas H. Assessment of right atrium volume by conventional CT or MR techniques: Which modality resembles in vivo reality?. European journal of radiology. 2016. DOI: 10.1016/j.ejrad.2016.03.005.

Предметные рубрики MeSH