Glucose measurements with accu check inform II versus hexokinase plasma method during surgery under general anesthesia, an observational cohort study.

Kaufmann Thomas, Slingerland Robbert J, Edens Mireille A, Olthof Cornelis G

Наблюдательное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Limited research exists on translation of in-vitro glucose measurement interfering compounds to the in-vivo situation. We investigated whether Point-of-Care glucose measurements by Accu Chek Inform II (ACI II) were accurate to monitor glucose concentrations during surgery with general anesthesia by comparing with the reference laboratory hexokinase plasma glucose test.

METHOD: Patients undergoing surgery with general anesthesia were included. Anesthesia was maintained with either Sevoflurane or Total intravenous anesthesia (TIVA). Prior to and after induction, blood glucose was measured with ACI II and the hexokinase test. Bland-Altman analysis was performed to assess method agreement. Subgroup analyses on glucose measurement differences per type of maintenance anesthesia were performed.

RESULTS

Thirty-nine patients were included, and 78 measurements were performed. All paired measurements had clinically acceptable agreement with a percentage error of 10.0% (95% CI 8.0 to 11.9). The mean difference (95% limits of agreement) between ACI II and hexokinase for all measurements was 0.0 mmol/L (-0.7 to 0.7 mmol/L). Before induction (n = 39), mean difference was -0.1 mmol/L (-0.6 to 0.4 mmol/L), and after induction (n = 39), mean difference was 0.1 mmol/L (-0.8 to 0.9 mmol/L). Further investigation showed the difference varied per test for patients receiving Sevoflurane compared to patients receiving TIVA (-0.2 ± 0.4 mmol/L vs. 0.4 ± 0.3 mmol/L, p < 0.001). Before and after induction, the difference between ACI II and hexokinase measurements increased for patients receiving Sevoflurane compared to patients receiving TIVA (0.4 ± 0.4 mmol/L vs. -0.4 ± 0.3 mmol/L, p < 0.001).

CONCLUSION

The agreement between glucose measurements using ACI II and the reference laboratory hexokinase test was clinically acceptable with a percentage error of 10.0% (95% CI 8.0 to 11.9). The use of TIVA may negatively affect the measurement performance of the ACI II.

Для цитирования: Kaufmann Thomas, Slingerland Robbert J, Edens Mireille A, Olthof Cornelis G. Glucose measurements with accu check inform II versus hexokinase plasma method during surgery under general anesthesia, an observational cohort study.. Journal of clinical monitoring and computing. 2023. DOI: 10.1007/s10877-023-01072-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.