Ex Vivo Measurement of Human Blood T1 at 5 T: Dependence on Hematocrit and Oxygenation.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Yu Shuwan, Dong Yuan, Gu Yuxiao, Yang Ne, Ma Lele, Xu Qingyu, Ma Ping, Zhou Baoxin, Chen Shuo, Chen Yonglin, Jiang Dengrong, Zhao Xihai, Lin Zixuan

Аннотация

PURPOSE: To determine human blood longitudinal relaxation time (T1) ex vivo at 5 T using modified Look-Locker inversion recovery (MOLLI), and to evaluate its dependence on hematocrit (Hct) and oxygenation (Y).

METHODS: A MOLLI 5(3)3 protocol was simulated across T1 values of 1000-3000 ms to optimize the sampling interval. Aqueous gadolinium phantoms were scanned using optimized MOLLI and multi-inversion-time recovery spin echo (IR-SE) acquisitions to derive a calibration equation. Arm venous blood from 27 healthy volunteers (mean age: 23.7 ± 4.3 years; 10 males) was measured using optimized MOLLI, with three repetitions per sample. Hct and Y were measured by blood gas analysis, and their relationships with calibrated T1 were evaluated.

RESULTS: The minimum root mean square error (RMSE) was obtained at a MOLLI sampling interval of 2000 ms. In phantoms, MOLLI and IR-SE T1 showed a strong linear relationship: T1,IR-SE = 1.185 × T1,MOLLI-93.746 (R2 = 0.998, p < 0.0001). In blood, MOLLI showed high repeatability, with an overall repeatability RMSE of 1.80 ms and a mean coefficient of variation of 0.10% ± 0.08%. Calibrated blood T1 was 1933.6 ± 114.2 ms, range 1727.0-2128.0 ms. Blood 1/T1 dependence on Hct and Y: 1/T1 = 0.447-0.193 × Y + 0.447 × Hct + 0.150 (Y × Hct) (R2 = 0.855, p < 0.0001), yielding an arterial blood T1 of 1971 ms at Y = 0.98 and Hct = 0.42.

CONCLUSION: Optimized and phantom calibrated MOLLI enabled rapid measurement of human ex vivo blood T1 at 5 T. Blood T1 at 5 T was higher than commonly reported adult values at 3 T and varied with both Hct and Y.