MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/0218492312438741

Медицинская библиотека · Asian cardiovascular & thoracic annals · 2012

A new concept in the sewing rings for mechanical heart valves.

Kurian Mathew V, Umasankar Payanam R, Sabareeswaran Arumugam, Shenoy Sachin J, Subban Vijayakumar, Joseph Mathew K

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Valve-associated complications challenge the quality of life and longevity after heart valve replacement. Inappropriate healing may be a contributing factor.

OBJECTIVE

To evaluate a new design mechanical heart valve in an animal model.

METHODS AND

RESULTS

The new valve was fabricated by substituting the sewing ring with an inbuilt suture ridge and an overlying fabric flap. It improved the effective orifice area. Animal experiments were performed on pigs to compare this valve with standard valve models. The animals were kept on dual antiplatelet drugs. Six of the 8 test animals survived the observation period of 140 days compared to 2 of the 6 controls. Among the test valves, one had thrombosis and 3 had significant tissue hyperplasia, whereas 5 control valves had thrombosis associated with significant tissue hyperplasia. Three test valves had paravalvular defects compared to none in the control group. Histology showed good tissue incorporation of the fabric flap of the test valves, whereas the control valves had tissue infiltration limited to the peripheral fabric layer of the sewing ring.

CONCLUSION

The new valve has improved effective orifice area, and the animal study showed better survival, good healing, and a lower incidence of thrombosis and tissue hyperplasia.

Для цитирования: Kurian Mathew V, Umasankar Payanam R, Sabareeswaran Arumugam, Shenoy Sachin J, Subban Vijayakumar, Joseph Mathew K. A new concept in the sewing rings for mechanical heart valves.. Asian cardiovascular & thoracic annals. 2012. DOI: 10.1177/0218492312438741.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.