DOI 10.1093/europace/euaa026

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2020

Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator.

Rudic Boris, Tülümen Erol, Fastenrath Fabian, Röger Susanne, Goranova Diana, Akin Ibrahim, Borggrefe Martin, Kuschyk Jürgen

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: Inappropriate shocks (IAS) remain a challenge for patients and physicians after implantation of the subcutaneous implantable cardioverter-defibrillator (S-ICD). The aims were to assess and characterize different patterns of IAS.

METHODS AND

RESULTS

Two hundred and thirty-nine patients were implanted with an S-ICD between 2010 and 2018 for primary and secondary prevention. Follow-up data of at least 6 months were analysed. During a mean follow-up of 34.9 ± 16.0 months, a total of 73 shocks occurred in 38 patients (6%). Forty-three (59%) shocks were considered appropriate due to ventricular tachycardia/ventricular fibrillation, while 30 (41%) were inappropriate and occurred in 19 patients (8%). Myopotentials/noise was the most frequent cause of inappropriate shocks (n = 8), followed by T-wave oversensing (n = 6) and undersensing of the QRS, resulting in adaptation of the automatic gain control and inappropriate shock (n = 5). Seventy-four percent of all IAS occurred on the primary vector, while no IAS occurred on the alternate vector. In seven of eight patients (88%), IAS related to myopotentials have occurred on the primary sensing vector. Multivariate analysis identified taller patients, primary sensing vector and first-generation S-ICD device as predictors for IAS. SMART pass effectively reduced the occurrence of IAS in the second-generation S-ICD system.

CONCLUSION

Inappropriate therapies are less frequently observed on the alternate vector. The primary vector seems to be unfavourable with regard to oversensing caused by myopotentials. Inappropriate shocks were associated with an increased rate of rehospitalization but not mortality. These observations have implications for the prevention of inappropriate S-ICD shocks.

Для цитирования: Rudic Boris, Tülümen Erol, Fastenrath Fabian, Röger Susanne, Goranova Diana, Akin Ibrahim, Borggrefe Martin, Kuschyk Jürgen. Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2020. DOI: 10.1093/europace/euaa026.

Предметные рубрики MeSH