Metallic residues as a source of artifacts in magnetic resonance imaging of the temporomandibular joint.

Tetsumura A, Honda E, Sasaki T, Kino K

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To establish the source of metallic artifacts on postoperative MRI of the temporomandibular joint after arthroplasty.

METHODS

An experiment was carried out to evaluate MR artifacts which were thought to have been caused by minute metallic particles deposited at surgery. Four types of cutting instruments were used to cut an extracted tooth: a diamond bur attached to a dental air turbine handpiece, a steel bur and bone bur attached to a dental micromotor handpiece and a bone file. The tooth fragments were collected in plastic centrifuge tubes in water. These samples were evaluated with six MR sequences.

RESULTS

MR artifacts were clearly demonstrated in samples using the bone bur and the bone file but not the diamond and steel burs.

CONCLUSIONS

The artifacts seen on postoperative MRI were derived from minute metal particles from the bone bur or file used for condylar arthroplasty. Metal artifacts should be considered when interpreting postoperative MR images.

Для цитирования: Tetsumura A, Honda E, Sasaki T, Kino K. Metallic residues as a source of artifacts in magnetic resonance imaging of the temporomandibular joint.. Dento maxillo facial radiology. 1999. DOI: 10.1038/sj/dmfr/4600429.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.