Effect of motion artifacts and scan circle displacements on Cirrus HD-OCT retinal nerve fiber layer thickness measurements.

Taibbi Giovanni, Peterson Gregory C, Syed Misha F, Vizzeri Gianmarco

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the effect of scan circle displacements on retinal nerve fiber layer thickness (RNFLT) measurements in Cirrus HD-OCT scans with motion artifacts affecting the optic disc.

METHODS

In this cross-sectional study, 70 scans from 18 healthy eyes and 100 scans from 26 glaucomatous eyes were divided into 85 pairs, each composed by a scan with one motion artifact affecting the optic disc, and a scan from the same eye without motion artifacts. En face images underwent automated realignment, and horizontal/vertical scan circle displacements were determined. Multiple regression analysis evaluated the relationship between scan circle displacements and RNFLT change.

RESULTS

Scans with motion artifacts showed similar displacements in healthy and glaucomatous eyes (P values ≥ 0.08). Average RNFLT and quadrants were relatively unchanged, while clock-hours showed more changes (e.g., in glaucomatous eyes, clock-hour-7 RNFLT was lower in scans with motion artifacts, P = 0.05). Scan circle displacements produced average RNFLT changes above test-retest variability in 3/85 cases (3.53%). Retinal nerve fiber layer thickness tended to decrease in sectors moved away from the disc and to increase in sectors closer to the disc (R(2) ≤ 0.40 and R(2) ≤ 0.22 in healthy and glaucomatous eyes, respectively). In healthy eyes, horizontal displacements ≥ 423 and 325 μm were associated with average and quadrant RNFLT changes above test-retest variability, respectively.

CONCLUSIONS

Scan circle displacements occurred in all scans with motions artifacts affecting the optic disc. Average RNFLT and quadrants were more robust than clock-hours. Because motion artifacts may be difficult to detect, clinicians should carefully inspect en face OCT images for their presence and interpret clock-hour results cautiously.

Для цитирования: Taibbi Giovanni, Peterson Gregory C, Syed Misha F, Vizzeri Gianmarco. Effect of motion artifacts and scan circle displacements on Cirrus HD-OCT retinal nerve fiber layer thickness measurements.. Investigative ophthalmology & visual science. 2014. DOI: 10.1167/iovs.13-13276.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.