DOI 10.1111/j.1464-410x.2011.10510.x

Медицинская библиотека · BJU international · 2011

Safer urethral catheters: how study of catheter balloon pressure and force can guide design.

Wu Alex K, Blaschko Sarah D, Garcia Maurice, McAninch Jack W, Aaronson David S

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To better define urethral catheter balloon pressures and extraction forces during traumatic placement and removal of urethral catheters. To help guide design for safer urethral catheters.

MATERIALS AND

METHODS

Measurements of balloon pressure were made upon filling within the urethra vs the bladder. Extraction forces were measured upon removal of a catheter with a filled balloon from the bladder. Models for the bladder and urethra included an ex vivo model (funnel, 'bladder', attached to a 30 F tube, 'urethra') and fresh human male cadavers. The mean (SEM) balloon pressures and extraction forces were calculated.

RESULTS

In the ex vivo model, the mean (SEM) pressures upon filling the balloon with 10 mL were on average three-times higher within the ex vivo'urethra' (177 [6] kPa) vs 'bladder' (59 [2] kPa) across multiple catheter types. In the human cadaver, the mean balloon pressure was 1.9-times higher within the urethra (139 [11] kPa) vs bladder (68 [4] kPa). Balloon pressure increased non-linearly during intraurethral filling of both models, resulting in either balloon rupture (silicone catheters) or 'ballooning' of the neck of the balloon filling port (latex catheters). Removal of a filled balloon per the ex vivo model 'urethra' and cadaveric urethra, similarly required increasing force with greater balloon fill volumes (e.g. 9.34 [0.44] N for 5 mL vs 41.37 [8.01] N for 10 mL balloon volume).

CONCLUSIONS

Iatrogenic complications from improper urethral catheter use is common. Catheter balloon pressures and manual extraction forces associated with urethral injury are significantly greater than those found with normal use. The differences in pressure and force may be incorporated into a safer urethral catheter design, which may significantly reduce iatrogenic urethral injury associated with catheterization.

Для цитирования: Wu Alex K, Blaschko Sarah D, Garcia Maurice, McAninch Jack W, Aaronson David S. Safer urethral catheters: how study of catheter balloon pressure and force can guide design.. BJU international. 2011. DOI: 10.1111/j.1464-410x.2011.10510.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.